Pump Installation
5.5
Installing the pH/ORP Electrode
Note: Improve stability in
high purity water by
grounding the solution
near the electrode.
Different mounting equipment is available for the three configurations for mounting the
electrode. The electrode can be configured as an in-line unit installed in a full pipeline (see
Figure 15) or on the end of a dip tube as a submersible unit in a tank or other vessel. The
electrode also can be placed in a flow-thru cell for applications of flow with very little
pressure (see Figure 16). Maximum pressure is 35 psi for in-line installation or 40 feet of
immersion depth for submersible installation. Do not deviate orientation more than
90 degrees from vertical for the three types of installation. The configuration used
determines whether the bushing or collar is threaded with the reference tube and cable
prior to installation. Maximum temperature for the electrode is 60 °C.
Summary of Contents for EC1000
Page 2: ...2...
Page 8: ...8...
Page 12: ...12...
Page 14: ...14...
Page 16: ...16...
Page 18: ...18 Figure 1 Controller Front Panel...
Page 20: ...20 Figure 2 Pump Module...
Page 24: ...24 Figure 6 pH Control Loop...
Page 54: ...54...
Page 56: ...56...
Page 58: ...58 Figure 12 Controller Installation Panel Mount...
Page 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Page 70: ...70...
Page 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Page 74: ...74...
Page 80: ...80...
Page 83: ...83 Figure 24 Sensor Replacement...
Page 85: ...85 Figure 25 Hach One pH Process Electrode...
Page 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Page 102: ...102...
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 108: ...108...
Page 112: ...112...
Page 116: ...116...
Page 122: ...122...
Page 123: ...123...
Page 124: ......