Figure 13 Get data log
Figure 14 Data logger
6.3 Set back the instrument to factory settings
1.
Install the service cable VAA10036.99.G00 in the DR300. Refer to
2.
Start the VAP235 software.
3.
Connect the service cable VAA10036.99.G00 to an USB port.
4.
Push on the DR300.
5.
Push
Connect
. The display of the DR300 shows serial.
6.
First save the customer data and get a data log. Refer to
on page 61.
7.
Push
Set delivery state
. Refer to
. The data log, event log and the user
calibration will be deleted.