Figure 7 General settings
Preparation for inspection
1.
Install the service cable VAA10036.99.G00 in the DR300. Refer to
2.
Connect the VAA10036.99.G00 to an USB port.
3.
Select the correct com port in the drop-down menu near by the Connect button. Refer
to
. Go to windows device manager and look for USB Serial Port (COMX).
Plug in and out the USB to see which com port the USB has.
4.
Push on the DR300.
5.
Push
Connect
.
6.
Push
Start inspection
.
7.
Enter the set number of the
VAA10038.99.V00
. The set number is shown and
automatically used for further inspections.
8.
Follow the instructions to complete the first inspection. Refer to
the test filter set, service cable, PC and VAP235 software
on page 21.
Inspection