Problem
Possible cause
Solution
The instrument
reading is
different from
the online
analyzer
reading.
The sample is not a
representative sample
and is not well mixed.
Make sure to get a representative
sample. Refer to the instructions in
the
Sample Collection
section of the
test procedure.
The instrument
reading is
different from
the online
analyzer
reading.
The grab sample was
not analyzed
immediately.
Note: The DPD method is
less accurate in the lower
ppb range compared to the
fluorescence method.
Other oxidants do not
interfere with the
fluorescence method.
Analyze the grab samples
immediately. The chlorine
concentration decreases quickly in
ultra low range chlorine samples.
Section 12 Consumables and replacement parts
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause
personal injury, damage to the instrument or equipment
malfunction. The replacement parts in this section are approved by
the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact
the appropriate distributor or refer to the company website for contact
information.
Consumables
Description
Item no.
DR1300 FL Fluorescence Standards Kit
34630000
Reagent kit, ULR Fluorescence Total Chlorine,
includes:
Sample vial holder, two16-mm vials with caps and two
reagent bottles in dropper bottles
34252000
Reagent kit, ULR Fluorescence Free Chlorine, includes:
Sample vial holder, two16-mm vials with caps and two
reagent bottles in dropper bottles
34251000
English
25
Summary of Contents for DR1300 FL
Page 2: ...Table of Contents English 3 Espa ol 27 Fran ais 54 Portugu s 81 108 129 153 2...
Page 112: ...1 DR1300 FL 1 3 3 2 4 LCD 3 112...
Page 113: ...3 4 2 2 1 DR1300 FL 4 2 5 USB 4 3 2 6 AA 4 113...
Page 114: ...4 AA 3 3 114...
Page 115: ...5 5 1 4 4 1 Enter 5 4 Right 2 Left 5 Down 3 Up 115...
Page 116: ...5 2 5 5 1 5 2 6 3 7 4 116...
Page 118: ...7 MSDS SDS 119 1 Enter 2 3 a b c d 119 e f g 4 118...
Page 124: ...10 10 1 AA 114 10 2 10 3 124...
Page 125: ...11 120 120 120 120 125...
Page 126: ...124 120 120 120 DR1300 FL DPD ppb 12 126...
Page 128: ...1 3563500 16 mm 6 pkg 100866 128...
Page 131: ...3 1 3 1 1 131...
Page 133: ...1 2 3 3 2 Hach DR1300 FL DR1300 FL DR1300 FL DR1300 FL 3 3 2 DR1300 FL 1 DR1300 FL 133...
Page 134: ...1 DR1300 FL 1 3 3 2 4 LCD 3 134...
Page 135: ...3 4 2 2 1 DR1300 FL 4 2 5 Bluetooth USB 4 3 2 6 AA 4 4 Bluetooth 135...
Page 136: ...4 AA 4 3 3 136...
Page 137: ...5 5 1 4 4 1 Enter 5 4 2 5 3 137...
Page 138: ...5 2 5 5 1 SYSTEM 5 2 TESTS 6 CALIBRATION 3 7 POWER 4 138...
Page 140: ...Update Program Factory Default Factory Exit 7 MSDS SDS 141 1 Enter 2 TESTS 140...
Page 141: ...3 a b TIMER START c d 141 e f g BACK READ 4 7 1 7 2 1 http www hach com 2 DR1300 FL 3 141...
Page 147: ...LOW BATTERY 4 AA 136 10 2 10 3 147...
Page 148: ...11 SYSTEM Diagnosis Sample Collection 142 143 148...
Page 149: ...142 142 147 142 142 143 149...
Page 150: ...DR1300 FL Sample Collection DPD ppb 12 150...
Page 152: ...1 3563500 16mm 6 100866 152...
Page 155: ...3 1 1 3 1 2 3 1 3 ICES 003 155...
Page 157: ...DR1300 FL 1 DR1300 FL 1 DR1300 FL 1 3 3 Keypad 2 4 LCD 3 157...
Page 158: ...3 4 2 2 1 DR1300 FL 4 2 5 Bluetooth USB4 3 2x 6 AA 4 Bluetooth 158...
Page 159: ...4 AA 3 3 159...
Page 160: ...5 5 1 Keypad 4 4 1 Enter 5 4 2 5 3 160...
Page 161: ...5 2 Display 5 5 Main menu 1 SYSTEM 5 2 TESTS 6 CALIBRATION 3 7 POWER 4 161...
Page 163: ...Factory Default Factory Exit 7 MSDS SDS 164 1 Enter 2 TESTS 3 a 163...
Page 165: ...8 MSDS SDS 8 1 1 Enter 2 3 164 4 CALIBRATION 5 CAL RB 6 a b c d e Enter 7 Enter 8 165...
Page 170: ...11 SYSTEM Diagnosis 165 166 165 170...
Page 171: ...165 169 165 165 166 DR1300 FL DPD 171...
Page 174: ......
Page 175: ......