Expanded programs
6.7.4
AQA - Analytical Quality Assurance
1.
Press
AQA
in the System Checks menu to configure AQA
measurement or
AQA
in the toolbar to the right of the
display, to carry out AQA measurement.
The AQA menu in System Checks contains the program for
checking the analytical quality. Configuration is done here,
for example the individual working method being controlled by
standard solutions or the cross-sensitivities of the analyses
identified by standard additon solutions, multiple determinations
and dilutions.
As optional accessories for full individual control, standard and
top-up solutions are available (
6.7.4.1
AQA Standard configuration
The working method, photometers, and accessories can be
checked by the analysis of a standard solution.
1.
Press
AQA Standard
A screen with standard sets opens (not available on all
models), alternatively an individual standard can be defined.
2.
Select
Options > New
to define a new standard.
Note: If a standard has not yet been defined, in some models the
alphanumeric keypad is displayed directly.
3.
Enter the
name
of the standard solution and confirm with
OK
.
4.
Enter the
Concentration
and
select the
Unit
by pressing the
corresponding fields. Enter the
Confidence Interval
and
select the
Unit
by pressing the corresponding fields.
5.
Press
OK
to confirm the entry.
6.
All the standard solutions entered are shown on the screen.
7.
Select
Options > Edit
to alter the settings.