Most laboratory detergents are used at recommended concentrations. Neutral detergents, such as
Liquinox, are safer to use when regular cleaning is necessary. To decrease the cleaning times,
increase the temperature or use an ultrasonic bath. To complete the cleaning, rinse a few times with
deionized water and then let the sample cell air dry.
Sample cells may also be cleaned with acid, followed by a thorough rinse with deionized water.
Note: Always use acid to clean sample cells that were used for low-level metal tests.
Special cleaning methods are necessary for individual procedures. When a brush is used to clean
sample cells, take extra care to avoid scratches on the interior surfaces of the sample cells.
Replace the batteries
W A R N I N G
Explosion hazard. Expired batteries can cause hydrogen gas buildup inside the instrument. Replace
the batteries before they expire and do not store the instrument for long periods with the batteries
installed.
on page 7 for battery replacement.
Troubleshooting
Problem
Possible cause
Solution
Absorbance > 3.5!
The measured absorbance is more than 3.5.
Dilute the sample and repeat the
reading.
Over range!
The concentration is more than the upper limit
of the current method.
Dilute the sample and repeat the
reading.
Under range!
The concentration is less than the lower limit of
the current method.
Examine the sample.
Reading Failed!
There is an electronic or optical defect.
Contact technical support.
Initializing Failed!
While initializing, the instrument failed.
Set the instrument to off and then
to on.
Contact technical support.
Calibration Needed!
The calibration data is corrupt.
Contact technical support.
Air Scan Needed!
The calibration data is corrupt.
Contact technical support.
Please attach the
module!
The module is not found when data is sent.
Insert a module.
Failed in sending data!
Make sure that the module is
inserted and latched correctly.
English
15
Summary of Contents for DR 1900
Page 2: ...English 3 Fran ais 16 Espa ol 30 Portugu s 44 58 71 85 98 2...
Page 59: ...10 x 10 mm 1 in 13 mm 16 mm 1 in 1 cm 10 mL III CE 1 EU 2 1 59...
Page 61: ...1 1 4 2 5 3 2 2 1 DR 1900 4 2 4x 5 1 10 mL 3 AA 4 6 61...
Page 62: ...4 AA 4 3 3 1 4 1 1 A 13 10 B 1 C 1 1 cm 10 mL D 16 62...
Page 63: ...4 5 5 6 63...
Page 64: ...6 1 4 2 5 3 6 7 64...
Page 65: ...7 1 8 2 9 OK 3 10 4 11 5 12 6 13 7 TNTplus 1 1 POWER POWER 1 Auto Shutoff 1 2 65...
Page 69: ...MSDS SDS Liquinox 62 3 5 3 5 69...
Page 70: ...70...
Page 74: ...1 1 4 2 5 3 2 2 1 DR 1900 4 2 4 5 1 1 10 mL 3 4 6 74...
Page 75: ...3 4 NiMH 4 3 3 1 1 4 1 1 A 13 mm 10 mm B 1 C 1 1 cm 10 mL D 16 mm 75...
Page 76: ...4 5 5 6 76...
Page 77: ...6 1 4 2 5 3 6 ID ID 7 77...
Page 83: ...MSDS SDS Liquinox 75 Absorbance 3 5 3 5 3 5 Over range Under range Reading Failed 83...
Page 86: ...500 ID GLP ID 10 x 10 mm 1 13 mm 16 mm 1 1 cm 10 mL Class II III CE 1 EU 2 1 86...
Page 88: ...1 1 4 2 5 3 88 2 2 1 DR 1900 4 2 4x 5 10 mL 1 3 AA 4 6 88...
Page 89: ...AA 4 NiMH 4 3 3 1 4 1 1 A 13 mm 10 mm B 1 C 1 1 cm 10 mL D 16 mm 89...
Page 90: ...4 5 5 6 90...
Page 91: ...6 1 4 2 5 3 6 ID ID 7 91...
Page 92: ...7 1 8 2 9 3 10 4 11 5 12 ID 6 AC 13 ID 7 TNTplus 1 1 1 92...
Page 94: ...1 2 03 00 ID ID 96 ID 96 ID ID 95 ID 95 T Abs Conc 94 94 1 2 2 2 94...
Page 96: ...ID ID ID 1 ID 2 ID ID ID ID ID ID ID 50 ID ID MSDS SDS Liquinox 96...
Page 97: ...89 89 3 5 3 5 97...
Page 101: ...DR 1900 340 800 nm AA LCK TNTplus 1 1 1 4 2 5 3 2 2 1 DR 1900 4 2 4x 5 1 1 10 mL 3 AA 4 6 101...
Page 102: ...AA NIMH 3 3 102...
Page 103: ...1 4 1 1 A 13 10 B 1 C 1 1 10 D 16 4 5 103...
Page 104: ...5 6 6 1 4 2 SETTINGS 5 3 6 HOME ID ID 7 104...
Page 110: ...102 3 5 3 5 Optical 110...
Page 111: ......