![Hach CM130 User Manual Download Page 73](http://html2.mh-extra.com/html/hach/cm130/cm130_user-manual_3212248073.webp)
Table 13 Alarms (red
) (continued)
Message
Description
Solution
Reboot the
device
A software error
occurred.
Measurements cannot
be done.
Push
menu
, then select Power > shut down. Wait for the analyzer
to shut down. When the display is off, disconnect the power cord
from the electrical outlet. Wait 1 minute. Connect the power cord to
the electrical outlet.
Remote Indicator
not detected!
The remote indicator is
not connected or there
is a problem with the
remote indicator.
1.
Make sure that the remote indicator cable is connected to the
analyzer and the remote indicator correctly. Refer to
the remote indicator cable to the remote indicator
on page 33
Connect the remote indicator cable to the analyzer
on page 40.
2.
If the problem continues, replace the remote indicator. Refer to
the instructions for use supplied with the new remote indicator.
3.
If the problem continues after the remote indictor is replaced, it
may be necessary to replace the analyzer. Contact technical
support.
Remote Indicator
error
The remote indicator
LEDs are not
functioning.
Speaker test
error
There is a problem with
the remote indicator
speaker.
Stir system error! The magnetic stirrer in
the colorimeter is not
detected.
1.
Look through the colorimeter window to see if there is a
magnetic stirrer. Refer to
on page 9.
2.
Make sure that the stir motor cable is connected to the front
panel. Refer to
Analyzer electrical connectors
on page 38.
3.
Make sure that nothing prevents the stir motor operation.
4.
If the problem continues, remove power from the analyzer.
Refer to
on page 61. Refer to the clinic
operating procedures.
5.
Replace the stir motor. Refer to the instructions for use supplied
with the new stir motor.
Software error!
A software error
occurred.
Push
menu
, then select Power > shut down. Wait for the analyzer
to shut down. When the display is off, disconnect the power cord
from the electrical outlet. Wait 1 minute. Connect the power cord to
the electrical outlet.
Table 14 Miscellaneous (no notification message)
Problem
Possible cause
Solution
The black and red bottle
caps were installed on the
wrong reagent bottles.
The reagent bottles were
installed in the wrong
positions in the analyzer.
1.
Remove the bottle caps from the reagent bottles.
2.
Use a clean, no-lint cloth to remove the reagent from
the exterior surface of the tubes.
3.
Install the black and red bottle caps on the correct
reagent bottles.
4.
Push
menu
, then select Maintenance > Prime
reagents.
5.
Do step 4 again.
The remote indicator
makes a continuous
sound.
The battery life of the
remote indicator is low.
Replace the battery when a continuous sound is heard.
Refer to
Replace the remote indicator battery
on page 60.
English
71
Summary of Contents for CM130
Page 1: ...DOC023 53 80540 CM130 11 2018 Edition 5 User Manual...
Page 2: ......
Page 18: ...Figure 9 Analyzer dimensions 16 English...
Page 65: ...Figure 22 Remove the pump clamp English 63...
Page 77: ...English 75...
Page 79: ...English 77...
Page 80: ...78 English...
Page 81: ...English 79...
Page 82: ...80 English...
Page 83: ...English 81...
Page 84: ...82 English...
Page 85: ...English 83...
Page 86: ...84 English...
Page 88: ...86 English...
Page 90: ...88 English...
Page 92: ...90 English...
Page 94: ...92 English...
Page 95: ...English 93...
Page 96: ...94 English...
Page 98: ...96 English...
Page 99: ...English 97...
Page 100: ...98 English...
Page 101: ...English 99...
Page 102: ...100 English...
Page 103: ...English 101...
Page 104: ...102 English...
Page 105: ...English 103...
Page 106: ...104 English...
Page 107: ...English 105...
Page 111: ......