![Hach BioTector B3500s User Manual Download Page 31](http://html1.mh-extra.com/html/hach/biotector-b3500s/biotector-b3500s_user-manual_3203147031.webp)
Page 31
2.3.2 Span Calibration
S P A N C A L I B R A T I O N 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
1 < T O C S P A N A D J U S T 1 1 . 0 0
2 2 1 . 0 0
3 3 1 . 0 0
4 T I C S P A N A D J U S T 1 1 . 0 0
5 2 1 . 0 0
6 3 1 . 0 0
1 0 R U N S P A N C A L I B R A T I O N
1 1 R U N S P A N C H E C K
1 2
1 3 S P A N P R O G R A M 6
1 4 S P A N A V E R A G E 4
1 5 R A N G E 1
1 6 T O C C A L S T D 1 0 0 . 0 m g C / l
1 7 T O C C H E C K S T D 5 0 . 0 m g C / l
1 8 T I C C A L S T D 1 2 5 . 0 m g C / l
1 9 T I C C H E C K S T D 6 0 . 0 m g C / l
2 2
2 3 - - > S P A N P R O G R A M
Span Calibration menu allows the user to enter the Span Adjust values manually, to start the Span Calibration
and Span Check cycles and to program the number of span reactions, span operation range and the
concentrations of the standard solutions used. Above menu displays the parameters for the TIC & TOC
systems. In TC and VOC systems, the relevant parameters, identified below, are displayed in this menu.
1.-3. TOC Span Adjust.
This menu item allows the user to set the TOC span adjust factors manually by
entering the STANDARD solution used and the calibrated average reaction RESULT at each range (1, 2
and 3). When the STANDARD and RESULT values are entered, system automatically calculates the
corresponding span factors of each parameter for each range. In TC and VOC systems, this function is
named as TC Span Adjust. In order to manually set the Span Adjust factors:
First enter the concentration of the standard solution used.
S P A N A D J U S T 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
S T A N D A R D R E S U L T
1 < T O C S P A N A D J U S T 1 0 0 . 0
2 2 1 . 0 0
3 3 1 . 0 0
4 T I C S P A N A D J U S T 1 1 . 0 0
Next enter the average result.
S P A N A D J U S T 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
S T A N D A R D R E S U L T
1 < T O C S P A N A D J U S T 1 0 0 . 0 9 9 . 5
2 2 1 . 0 0
3 3 1 . 0 0
4 T I C S P A N A D J U S T 1 1 . 0 0
When the ENTER key is pressed again, the new span factor is automatically calculated. In order to set the
span adjust factors to 1.00, enter 0.0 values for both standard and result.
Summary of Contents for BioTector B3500s
Page 17: ...Page 17 Software Menu Diagram...
Page 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Page 63: ...Page 63...
Page 74: ...Page 74 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 156: ...Page 156 Section 11 System Replacement and Spare Parts...
Page 162: ...Page 162 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Page 165: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...