8.2 COMMISSIONING MENU
The Commissioning menus are used during the commissioning and startup of the analyzer. The functions in
the menus are used to program system site specific settings.
Commissioning Menu Diagram
8.2.1 Reaction Time
REACTION TIME 7m30s
REACTION TIME displays the total reaction time (in minutes and
seconds) for range 1, based on all programmed settings in
System Program, System Program 1 menu.
INTERVAL 0m
INTERVAL is the time (0 minutes by default) added between
each reaction. Interval can be programmed on site if frequent
analysis is not necessary. The longest interval time, which can
be programmed, is 1440 minutes (or 1 day). A programmed
interval time would reduce the reagent usage significantly.
When BioTector automatically extends the reaction time due to
high level of TIC and/or TOC in the sample, if any INTERVAL
time is programmed, the extended reaction time is taken out
from the interval time.
BioTector automatically generates an INTERVAL time when the
user programs any SAMPLER, FORWARD and/or REVERSE
times, which exceeds the maximum allowable time in the
Sample Pump menu below. The system determines the
maximum allowable time from the System Program 1 settings in
System Program menu.
TOTAL 7m30s
TOTAL displays the total reaction time including the interval time
if programmed.
Summary of Contents for BioTector B3500e
Page 17: ...Page 17 Software Menu Diagram...
Page 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Page 61: ...Page 61...
Page 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Page 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...