![Hach BioTector B3500e User Manual Download Page 115](http://html1.mh-extra.com/html/hach/biotector-b3500e/biotector-b3500e_user-manual_3202243115.webp)
Page 115
8.3.4.3
Span Program
SPAN PROGRAM 5
SPAN PROGRAM defines the number of reactions to be carried
out for Span Calibration and Span Check cycles, which are
activated by RUN SPAN CALIBRATION/CHECK functions in
Span Calibration menu. If the Span Calibration or Span Check
is programmed as YES in New Regents Program menu, START
NEW REAGENT CYCLE function in Install New Reagents menu
will also run the relevant span cycle.
The span calibration/check reactions are run at a single range
programmed by the RANGE below. System calculates and uses
the same Span Adjust values for other ranges as well.
SPAN AVERAGE 3
SPAN AVERAGE is the number of span reactions the system
averages to calculate the Span Adjust factor.
RANGE 1
RANGE (1 by default) determines the range at which all Span
Calibration and Span Check reactions are carried out.
TOC CAL STD 10.0mgC/l
TOC CALIBRATION STANDARD defines the TOC standard
solution concentration level (mgC/l) to be used in Span
Calibration reactions.
If TOC CALIBRATION STANDARD is programmed as 0.0mgC/l,
system ignores the Span Calibration results and omits any
related warnings defined in TOC BAND below.
TOC CHECK STD 10.0mgC/l
TOC CHECK STANDARD defines the TOC standard solution
concentration level (mgC/l) to be used in Span Check reactions.
If TOC CHECK STANDARD is programmed as 0.0mgC/l,
system ignores the Span Check results and omits any related
warnings defined in TOC BAND below.
TIC CAL STD 5.0mgC/l
TIC CALIBRATION STANDARD defines the TIC standard
solution concentration level (mgC/l) to be used in Span
Calibration reactions.
If TIC CALIBRATION STANDARD is programmed as 0.0mgC/l,
system ignores the Span Calibration results and omits any
related warnings defined in TIC BAND below.
In TC systems, TIC CALIBRATION STANDARD is not
displayed.
TIC CHECK STD 5.0mgC/l
TIC CHECK STANDARD defines the TIC standard solution
concentration level (mgC/l) to be used in Span Check reactions.
If TIC CHECK STANDARD is programmed as 0.0mgC/l, system
ignores the Span Check results and omits any related warnings
defined in TIC BAND below.
In TC systems, TIC CHECK STANDARD is not displayed.
Summary of Contents for BioTector B3500e
Page 17: ...Page 17 Software Menu Diagram...
Page 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Page 61: ...Page 61...
Page 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Page 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...