![Hach BioTector B3500e User Manual Download Page 112](http://html1.mh-extra.com/html/hach/biotector-b3500e/biotector-b3500e_user-manual_3202243112.webp)
Page 112
SYSTEM PROGRAM 2 TCe
SAMPLE IN 2p , P-V
SAMPLE IN defines the amount of sample (in pulses) to be
injected into the reactor in TC mode at range 2.
8.3.3 Calibration Data
DECIMAL PLACES 3
DECIMAL PLACES setting defines the number of decimal
places (0, 1, 2 or 3) system displays in the reaction results and
in the relevant system menus.
The default unit for reaction results is mg/l “milligrams per liter”.
Depending on application requirements and preferences, it is
possible to configure the system to display the results in ppm
“parts per million”, in µg/l “micrograms per liter” or in ppb “parts
per billion”. For further information, contact your local distributor
or the manufacturer.
TOC/TC CALIBRATION 1
TOC/TC CALIBRATION 2
TOC/TC CALIBRATION 3
TIC CALIBRATION 1
TIC CALIBRATION 2
TIC CALIBRATION 3
Calibration menus are used to install the calibration data and
thus to calibrate the BioTector in factory. TOC CALIBRATION 1,
2 and 3 shows the calibration curves for range 1, 2 and 3
respectively. In TC and VOC systems, the TOC CALIBRATION
1/2/3 menus are named as TC CALIBRATION 1/2/3.
The first column shows the number of calibration points, the
second point [%] shows the calibration factors automatically
calculated from the calibration data, the third column
STANDARD shows the standard solution concentrations and the
fourth column shows the un-calibrated result (mgC/l) measured
and calculated from the CO
2
analyzer readings.
System calibration curves should not be modified by the user on
site unless a recalibration of the system is required at a different
analysis range.
8.3.4 Sequence Program
8.3.4.1
Average Program
LOG AVERAGE NO
LOG AVERAGE function enables (YES) or disables (NO by
default) the 24-hour averaging of the reaction results in the
reaction archive.
The reaction results obtained from manual samples are not
included in the averaging calculations.
AVERAGE UPDATE 00:00
If the LOG AVERAGE above is enabled (YES), the system
calculates a 24 hours average of the reaction results and
displays it in the reaction archive for each stream at the time
programmed for AVERAGE UPDATE (00:00, HR:MIN, by
default).
Summary of Contents for BioTector B3500e
Page 17: ...Page 17 Software Menu Diagram...
Page 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Page 61: ...Page 61...
Page 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Page 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...