102
Plumbing and Connection Options
A.16 Option 10a plumbing and connections
Option 10a uses an sc analyzer with any type of sample
preparation that delivers a continuous sample stream that cannot
be pressurized. The waste of the analyzer is discharged into an
open drain.
Refer to
and the following instructions for Option 10a:
1.
Install the sample preparation unit.
2.
Feed the sample line from the sample preparation unit through
the analyzer (
, item 2). Use Seal Plug #3 to secure.
3.
Feed the drain tube through the analyzer (item 3). Use Seal
Plug #3 to secure.
Note:
Tubes can be pushed through prepared holes on Seal Plug #3.
4.
Seal any unused holes with Seal Plug #3.
5.
Connect the drain tube to the T-fitting.
6.
Connect the sample line from the sample preparation to the
bottom inlet on the overflow vessel using the fittings.