7.2.2 Graphical display
The graph shows concentration and temperature measurements for each channel in use. The graph
supplies easy monitoring of trends and shows changes in the process.
1.
From the graphical display screen use the up and down arrow keys to select a graph and push
the
HOME
key.
2.
Select an option:
Option
Description
MEASUREMENT VALUE
Set the measurement value for the selected channel. Select
between Auto Scale and Manually Scale. For manual scaling enter
the minimum and maximum measurement values
DATE & TIME RANGE
Select the date and time range from the available options
Section 8 Operation
8.1 Configure the sensor
Use the CONFIGURE menu to enter identification information for the sensor and to change options
for data handling and storage.
1.
Push the
menu
key, select SENSOR SETUP>CONFIGURE.
2.
Use the arrow keys to select an option and push
enter
. To enter numbers, characters or
punctuation, push and hold the
up
or
down
arrow keys. Push the
right
arrow key to advance to
the next space.
Option
Description
EDIT NAME
Changes the name that corresponds to the sensor on the top of the
measure screen. The name is limited to 16 characters in any combination
of letters, numbers, spaces or punctuation.
SENSOR S/N
Allows the user to enter the serial number of the sensor, limited to
16 characters in any combination of letters, numbers, spaces or
punctuation.
MEAS UNITS
Changes the measurement units—Select the unit from the list available.
TEMP UNITS
Sets the temperature units to °C (default) or °F
FILTER
Sets a time constant to increase signal stability. The time constant
calculates the average value during a specified time—0 (no effect, default)
to 60 seconds (average of signal value for 60 seconds). The filter
increases the time for the sensor signal to respond to actual changes in
the process.
LOG SETUP
Sets the time interval for data storage in the data log—5, 30 seconds, 1, 2,
5, 10, 15 (default), 30, 60 minutes.
RESET
DEFAULTS
Sets the configuration menu to the default settings. All sensor information
is lost.
8.2 Calibration
8.2.1 About sensor calibration
The sensor characteristics slowly shift over time and cause the sensor to lose accuracy. The sensor
must be calibrated regularly to maintain accuracy. The calibration frequency varies with the
application and is best determined by experience.
20
English
Summary of Contents for 9586sc
Page 151: ...4 5 4 2 4 3 1 1 6 2 7 3 8 4 6 mm 4 9 6 8 mm 5 151...
Page 152: ...5 5 1 4 152...
Page 153: ...5 2 2 153...
Page 154: ...2 1 5 2 9 2 2 4 20 mA 2 6 10 2 3 7 2 4 Modbus Profibus HART 4 20 mA 8 2 5 2 1 5 2 2 24V 2 154...
Page 156: ...5 2 3 100 250 VAC 50 60 Hz 5A 250 VAC 5A 24 VDC 5A 156...
Page 158: ...158...
Page 159: ...5 2 5 NEMA IP NEMA 4X IP66 1 2 3 3 3 3 3 2 4 2 3 1 2 1 1 1 2 3 4 500 5 6 159...
Page 275: ...Suomi 275...
Page 336: ...4 1 4 1 1 4 1 2 336...
Page 337: ...EMC 4 1 3 IECS 003 A FCC 15 15 FCC 1 2 15 FCC 1 2 3 4 5 4 2 337...
Page 338: ...4 3 1 1 6 2 7 3 8 4 6 4 9 6 8 5 DIPA 338...
Page 339: ...5 5 1 4 339...
Page 340: ...5 2 2 340...
Page 341: ...2 1 5 2 9 2 2 4 20 2 6 10 2 3 7 2 4 Modbus Profibus HART 4 20 8 2 5 2 1 5 2 2 2 341...
Page 342: ...24 PE 100 240 24 100 240 24 18 12 AWG 80 C 176 F NEMA 4X IP66 3 18 24 24 15 20 1 2 342...
Page 343: ...1 1 L1 2 N PE 2 1 24 2 24 PE 343...
Page 344: ...5 2 3 100 250 50 60 5 100 250 5 24 5 5 2 4 5 5 NEMA IP NEMA 4X IP66 100 250 344...
Page 346: ...346...
Page 347: ...5 2 5 NEMA IP NEMA 4X IP66 1 2 3 3 3 3 3 2 4 2 3 1 2 1 1 1 2 3 4 500 5 6 347...
Page 349: ...3 4 6 Startup 333 1 2 3 4 5 7 7 1 4 349...
Page 350: ...4 1 5 BACK 2 SD 6 MENU 3 HOME 7 4 ENTER SD 7 2 5 350...
Page 382: ......
Page 383: ......