![Hach 9586sc Basic User Manual Download Page 17](http://html1.mh-extra.com/html/hach/9586sc/9586sc_basic-user-manual_3141587017.webp)
5.4 Install the reagents
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.
1.
Fill the empty sample conditioning bottle with DIPA (diisopropylamine) and enter the hazardous
reagent information on the blank label attached to the bottle.
2.
Install the sample conditioning bottle on the front of the analyzer.
3.
Connect the supplied tube to the measuring cell. Install the tube completely into the bottle of
conditioning solution. Make sure that the tube fittings are tight to prevent leakage of fumes from
the conditioning bottle into the atmosphere.
4.
If the chemical zero cartridge is used, install the cartridge on the rear of the analyzer.
Section 6 Startup
Make sure that the flow rate and pressure do not exceed the values in
on page 3.
1.
Open the valve on the sample line to let sample flow through the analyzer.
2.
Turn the knob on the flow meter to set the flow rate.
3.
Examine the plumbing for leaks and stop any leaks if found.
4.
Apply power to the controller.
5.
Make the applicable menu selections when the controller starts.
Section 7 User interface and navigation
7.1 User interface
The keypad has four menu keys and four directional keys as shown in
.
English
17
Summary of Contents for 9586sc
Page 151: ...4 5 4 2 4 3 1 1 6 2 7 3 8 4 6 mm 4 9 6 8 mm 5 151...
Page 152: ...5 5 1 4 152...
Page 153: ...5 2 2 153...
Page 154: ...2 1 5 2 9 2 2 4 20 mA 2 6 10 2 3 7 2 4 Modbus Profibus HART 4 20 mA 8 2 5 2 1 5 2 2 24V 2 154...
Page 156: ...5 2 3 100 250 VAC 50 60 Hz 5A 250 VAC 5A 24 VDC 5A 156...
Page 158: ...158...
Page 159: ...5 2 5 NEMA IP NEMA 4X IP66 1 2 3 3 3 3 3 2 4 2 3 1 2 1 1 1 2 3 4 500 5 6 159...
Page 275: ...Suomi 275...
Page 336: ...4 1 4 1 1 4 1 2 336...
Page 337: ...EMC 4 1 3 IECS 003 A FCC 15 15 FCC 1 2 15 FCC 1 2 3 4 5 4 2 337...
Page 338: ...4 3 1 1 6 2 7 3 8 4 6 4 9 6 8 5 DIPA 338...
Page 339: ...5 5 1 4 339...
Page 340: ...5 2 2 340...
Page 341: ...2 1 5 2 9 2 2 4 20 2 6 10 2 3 7 2 4 Modbus Profibus HART 4 20 8 2 5 2 1 5 2 2 2 341...
Page 342: ...24 PE 100 240 24 100 240 24 18 12 AWG 80 C 176 F NEMA 4X IP66 3 18 24 24 15 20 1 2 342...
Page 343: ...1 1 L1 2 N PE 2 1 24 2 24 PE 343...
Page 344: ...5 2 3 100 250 50 60 5 100 250 5 24 5 5 2 4 5 5 NEMA IP NEMA 4X IP66 100 250 344...
Page 346: ...346...
Page 347: ...5 2 5 NEMA IP NEMA 4X IP66 1 2 3 3 3 3 3 2 4 2 3 1 2 1 1 1 2 3 4 500 5 6 347...
Page 349: ...3 4 6 Startup 333 1 2 3 4 5 7 7 1 4 349...
Page 350: ...4 1 5 BACK 2 SD 6 MENU 3 HOME 7 4 ENTER SD 7 2 5 350...
Page 382: ......
Page 383: ......