28
Programming the Instrument
5.3.1 Automatic Temperature Compensation
The sensor continuously measures the sample temperature. Concentration values
are automatically converted to a reference temperature of 25 °C. To set Automatic
Temperature Compensation:
1.
From the Main screen, press the
RIGHT ARROW
key.
2.
Use the
LEFT ARROW
key to highlight PROGRAMMING. Press
ENTER
.
3.
Use the
LEFT ARROW
key to highlight MEASURE. Press
ENTER
.
4.
Press
ENTER
to select TEMP.
COMP.
5.
In the TYPE field, select Auto with the
RIGHT ARROW
key.
6.
Press
ESC
to exit.
5.3.2 Manual Temperature Compensation
1.
From the Main display, press the
RIGHT ARROW
key.
2.
Use the
LEFT ARROW
key to highlight PROGRAMMING. Press
ENTER
.
3.
Use the
LEFT ARROW
key to highlight MEASURE. Press
ENTER
.
4.
Press
ENTER
to select
TEMP. COMP.
5.
In the TYPE field, select MANUAL with the
RIGHT ARROW
key.
6.
Enter the sample temperature. Press
ESC
when finished. Press
ENTER
until
the next menu option is highlighted.
5.4 Programming mA Outputs
The signals generated by the analog outputs allow measurement data to be
transmitted to any external control or recording device.
1.
From the Main display, press the
RIGHT ARROW
key.
2.
Use the
LEFT ARROW
key to highlight PROGRAMMING. Press
ENTER
.
3.
Use the
LEFT ARROW
key to highlight mA OUTPUTS. Press
ENTER
.
4.
Use the
LEFT ARROW
key to highlight OUTPUT 1, OUTPUT 2, SPECIAL
PROG, or TEST.
5.