18
Installation
Figure 12
Removing the Shielding Plate
Figure 13
Proper Wire Preparation and Insertion
4.4.4 Alarm Connections
DANGER
The relay connection area is
designed for only high voltage
(100–240 V ac) connections. A
shock hazard can exist if low
voltage (<30V) connections are
simultaneously made in the relay
connection area.
The product enclosure is IP65; however, achieving this protection level depends
on the use of proper hardware during installation. Use standard off-the-shelf
NEMA-approved conduit hardware. Take care to select the style of conduit
hardware that will seal to the enclosure wall when using flexible or rigid conduit.
The analyzer contains four unpowered alarm relays (3 A resistive,
100–230 V ac) designed for use with high voltage (greater than 30V-RMS). The
Normally Open and Common relay contacts are connected when an alarm
condition is active. The relay connector accepts 18–12 AWG wire. Determine wire
gauge based on load application. Less than 18 AWG is not recommended.
Note:
Current to the relay contacts
must be limited to 3 amps resistive.
Local removal of power from the
relays must be available in case of
emergency or for servicing the
product. If the relay contacts are
supplied with greater than 30 Volts,
power removal can be made with an
external switch and a 3-amp fuse or
with a switched 3-amp circuit
breaker.
1.
Remotely disconnect all instrument power before opening the analyzer door.
2.
Strip the insulation on each wire back ¼-inch.
3.
Remove the front cover. Remove the shielding plate inside the instrument.
4.
Route the wire through the back right conduit opening. Use the provided
cord-connected strain reliefs. Use appropriate NEMA sealing hardware.
5.
Insert the wire ends into the connector until the insulation seats against the
connector. Refer to
and
.
1.
Shielding Plate
3.
Ground stud (pulls the shield out of the instrument)
2.
Thumbscrew (loosen to remove shield)
1.
Strip ¼-inch of insulation.
2.
Seat insulation against connector with no bare wire exposed.
IC1
J1
D3
D4
RE4
RE3
D2
RE2
RE1
V1
V2
V3
V4
D1
C8
L2
C6
Z6
Model : 09125-A-2000
Serial No : 0104 549
Ind : D
Wiring located behind the guide plate shall be
rated at 80 C (176 F) or higher.
Vac : 100 - 240 V
Serial N°:
50/60Hz
25VA
Vac:
100 – 240
+
-
+
-
S1
S2
S3/Sys.Alarm
S4/Timer
Mains
PE
N
L
F1
T 2A L
250V
Caution: For continued protection
against risk of fire, replace only
with same type and rating fuse.
Analog
Outputs
Ser
ial com.
(option)
Rela
y
s
Amp
.
module
Preamp
.
Supply
I1
I2
RX/TX +
Temp +
RX/TX –
Shield
Temp –
Ref.
Counter
Work
GND
Aux
GND
+ V
GND
– V
Amperometry 9180
under shielding plate
K2
K9
K4
K5
IC2
J1
RP1
Ic4
K6
RP2
Ref
Temp +
Temp –
Counter
Work
Aux
9180
AMP
V1.03
P1
3
2
1
1
PE
Mains
N
L
2