Installation
Figure 9
Connecting ac Power with Customer-supplied Power Cord
Figure 10
Connecting ac Power with Customer-supplied Switch Box
4.4.2
Electrostatic Discharge (ESD) Considerations
Important Note:
The analyzer
circuit board components are
EXTREMELY sensitive to static
electricity damage in a powered or
unpowered state. Do not touch
circuit boards or its components
without wearing a grounded wrist
strap.
Delicate internal electronic components can be damaged by static electricity,
resulting in degraded instrument performance or eventual failure. Take the
following steps to prevent ESD damage to the instrument:
•
Before touching any instrument electronic components (such as printed circuit
cards and the components on them) discharge static electricity from your
body. This can be accomplished by touching an earth-grounded metal surface
such as the chassis of an instrument, or a metal conduit or pipe.
1.
Power Terminal
2.
Power Cord Strain Relief
1.
Power Terminal
2.
Conduit Strain Relief
IC1
J1
D3
D4
RE4
RE3
D2
RE2
RE1
V1
V2
V3
V4
D1
C8
L2
C6
Z6
Model : 09125-A-2000
Serial No : 0104 549
Ind : D
Wiring located behind the guide plate shall be
rated at 80 C (176 F) or higher.
Vac : 100 - 240 V
K2
K9
K4
K5
IC2
J1
RP1
Ic4
K6
RP2
Ref
Temp +
Temp –
Counter
Work
Aux
9180
AMP
V1.03
P1
1
2
IC1
J1
D3
D4
RE4
RE3
D2
RE2
RE1
V1
V2
V3
V4
D1
C8
L2
C6
Z6
Model : 09125-A-2000
Serial No : 0104 549
Ind : D
Wiring located behind the guide plate shall be
rated at 80 C (176 F) or higher.
Vac : 100 - 240 V
K2
K9
K4
K5
IC2
J1
RP1
Ic4
K6
RP2
Ref
Temp +
Temp –
Counter
Work
Aux
9180
AMP
V1.03
P1
1
2