![Hach 9186 Instrument Manual Download Page 13](http://html1.mh-extra.com/html/hach/9186/9186_instrument-manual_3179647013.webp)
Installation
13
4.3.3 Installing the In-line Filter
Note:
Rinse out sample system
before startup to prevent “crud
bursts” from clogging new filters.
1.
On the back of the panel, connect ¼ in. tubing to the elbow fitting on the top of
the measuring cell (see
).
2.
Connect the other end of the tubing to one end of the filter. Make sure that the
filter is oriented with the arrow pointing up.
3.
Connect another length of ¼ in. tubing to the bottom of the filter. Connect the
other end of the tubing to the elbow fitting on the right side of the pressure
regulator.
4.3.4 Connecting the Measuring Cell to the Transmitter
Insert the female end of the sensor connector into the 4-pronged connector at the
bottom of the measuring cell.
4.3.5 Choosing the Sample Line Location
Note:
Erratic readings will occur if a
sample is drawn from a location that
is too close to points of chemical
additions to the process stream, if
mixing is inadequate, or if the
chemical reaction is incomplete.
Selecting a good, representative sampling point is important for optimum
performance of your instrument. The sample analyzed must be representative of
the condition of the entire system.
Install sample line taps into the side of larger process pipes to minimize the
chance of ingesting sediment from the pipe line bottom or air bubbles from the top.
A tap projecting into the horizontal center of the pipe is ideal.
Figure 7
Sample Line Location
4.3.6 Connecting the Inlet Tubing
1.
Choose the sample line location as described above.
2.
Connect the ¼ in. O.D. tubing to the lower left elbow fitting on the Pressure
Regulator as shown in
.
4.3.7 Connecting the Outlet Tubing
On the back of the panel, connect the
3
/
8
-inch outlet tubing to the barbed fitting on
the back of the measuring cell. The tubing must not exceed 4 ft and it must drain
straight down.
Poor
Sediment (Typical)
Air (Typical)
Poor
Good
Best