Set up the reagents and standards
Make sure to set the sample flow rate and install the reagent bottles before this task is started.
1.
Push
menu
and then go to REAGENTS/STANDARDS.
2.
Select one of the options. Use the SET options when the volume of existing reagents/standards
needs to be adjusted. Use the RESET options when reagents/standards are replaced or renewed
(bottles are replaced).
Option
Description
SET REAGENT LEVEL
Sets the volume of a specific reagent in the reagent bottle to an estimated value.
Range: 1–100%.
SET STANDARD LEVEL
Sets the volume of a specific standard solution in the standard bottle to a
calculated approximate value. Range: 1–100%.
SET CLEANING LEVEL
Sets the volume of the cleaning solution in the cleaning solution bottle to an
estimated value. Range: 1–100%.
RESET REAGENT
LEVELS
Sets the volume of reagent in the reagent bottles to 100% full.
RESET STANDARD
LEVELS
Sets the volume of standard solution in the standard bottle to 100% full.
RESET CLEANING
SOLUTION LEVELS
Sets the volume of the cleaning solution in the cleaning solution bottle to 100%
full.
PRIME REAGENTS
Starts the flow of all reagents through the tube and valve system.
IMPORTANT:
prime the reagents each time the reagents are changed to remove bubbles
and push the new reagents through the system.
Grab sample options
The grab sample menu allows the user to analyze a specific sample or standard. Two options are
available:
• GRAB SAMPLE IN: This option is used to analyze an external sample or standard.
• GRAB SAMPLE OUT: This option is used to extract a sample directly from a sample line for
external analysis.
Measure a grab sample or standard
Use the grab sample funnel to measure grab samples taken from other areas in the system or to
measure a standard for a calibration verification.
1.
Collect 250–500 mL of the sample or standard in a clean container.
2.
Rinse the funnel outside of the unit with the collected sample.
3.
Install the funnel again.
4.
Select GRAB SAMPLE>GRAB SAMPLE IN.
5.
Complete the instructions on the display. The result shows for 5 minutes.
Note: To see the result after 5 minutes, go to the Event Log.
Get a grab sample from the analyzer
Use the grab sample tube to manually dispense a grab sample from one of the sample sources for
external analysis.
The sample source is measured by the analyzer immediately after the sample is dispensed. The
value of the sample source and the job identification number for the grab sample are shown on the
display.
8
English
Summary of Contents for 5500sc PO4 3-HR
Page 208: ...208 216 211 221 213 SD 224 213 224 214 1 1 1 6 2 7 3 8 4 9 5 Enter 2 3 208...
Page 209: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 4 0 100 1 209...
Page 210: ...45 75 10 45 50 75 10 50 2 6 1 2 3 ppb 48 SD 4 SD 210...
Page 211: ...4 SD 1 3 2 2 4 SD 1 1 1 2 3 4 211 4 1 2 3 4 2 211...
Page 212: ...5 6 7 0 14 bar 2 psi 0 28 bar 4 psi 55 mL 8 ENTER 9 10 224 216 216 1 enter enter 1 224 212...
Page 213: ...MSDS SDS 1 2 1 100 1 100 1 100 100 100 100 1 250 500 mL 2 213...
Page 214: ...3 4 5 5 5 1 2 3 4 211 1 2 3 9 10 240 15 ppb ppm mg L g L 1 5 1 16 10 214...
Page 215: ...10 216 223 216 4 20 mA 216 Modbus Profibus HART HACH55 215...
Page 218: ...PID 5 50 50 6 1 2 3 219 216 218...
Page 219: ...1 2 3 1 0 0 9 1 0 1 1 1 0 0 5 1 0 1 5 5 4 0 0 5 3 5 4 0 5 0 300 5 0 300 5 4 10 219...
Page 220: ...5 5 5 0 0 0 0 6 5 30 7 8 1 220...
Page 221: ...2 3 4 5 6 1 1 2 18 000 18 000 1 2 2 3 221...
Page 224: ...SD SD SD SD 2 GB 1 SD 4 211 2 SD SD SD 3 SD SD SD SD 1 SD 2 SD SD SD 3 4 SD 5 1 224...
Page 225: ...1 225 225 4 20 mA 1 2 3 ppb 4 1 2 ENTER 3 4 225...
Page 279: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 PROGNOSYS 4 PROGNOSYS PROGNOSYS 0 100 1 279...
Page 280: ...45 75 10 45 50 75 10 50 2 1 2 home 3 48 SD 4 SD 280...
Page 281: ...4 SD 1 3 2 2 4 SD 1 1 1 2 3 4 281 4 1 2 3 4 5 6 2 281...
Page 282: ...7 0 14 2 0 28 4 55 8 enter 9 10 294 286 286 1 enter enter 1 294 MSDS SDS 282...
Page 283: ...1 menu 2 1 100 1 100 1 100 100 100 100 1 250 500 2 3 4 5 5 5 283...
Page 284: ...1 2 3 4 281 1 2 menu 3 9 10 240 15 ppb ppm 1 5 1 16 10 10 284...
Page 285: ...285 292 286 4 20 286 Modbus Profibus HART HACH55 1 menu 2 285...
Page 287: ...4 20 1 menu 4 20 2 3 4 5 50 50 6 287...
Page 288: ...1 menu 2 3 288 286 1 menu 2 3 1 0 0 9 1 0 1 1 1 0 0 5 1 0 1 5 5 288...
Page 289: ...4 0 0 5 3 5 4 0 5 0 300 5 0 300 5 4 10 5 5 5 ONMAX 0 OFFMAX 0 ONMIN 0 min OFFMIN 0 min 6 289...
Page 290: ...5 30 7 8 1 2 3 4 5 6 1 1 menu 2 290...
Page 292: ...2 0 1023 511 3 A D REF A D A D 6 6 6 1 2 3 1 2 292...
Page 294: ...1 SD 2 SD SD SD 3 4 SD 5 1 1 cal 295 294 4 20 1 cal 2 3 ppb 4 294...
Page 295: ...1 cal 2 enter 3 4 295...
Page 370: ...4 SD 1 3 LED 2 2 4 SD 1 1 1 2 3 4 370 4 1 LANGUAGE 2 DATE FORMAT 3 DATE TIME 2 370...
Page 386: ...1 cal START MANUAL CAL 2 enter 3 YES NO 4 386...
Page 404: ......
Page 405: ......