Use an SD card
Use an SD memory card to update the software and firmware and to download the event and data
logs. The SD icon is visible in the upper status bar of the main measurement screen when a card is
installed. The manufacturer recommends to use an SD card with a minimum of 2 GB storage
capacity.
1.
Install the SD card (refer to
2.
Select SD CARD SETUP from the MAIN MENU.
Note: The SD CARD SETUP option shows only when an SD card is installed.
3.
Select an option.
Option
Description
UPGRADE SOFTWARE
(conditional)
Shows when an the upgrade file is available on the SD card. Select the specific
device for the upgrade.
SAVE LOGS
Selects the device for the download data and to save the logs for last day, last
week, last month or all.
MANAGE
CONFIGURATION
Saves and restores the backup settings, restore settings and/or transfer
settings between instruments.
WORK WITH DEVICES
READ DEVICE FILES—selects the data for each device to save on the SD
card. Options: sensor diag, measurement data (curve data for a measurement
cycle), cal history, cal data and/or test script. WRITE DEVICE FILES—shows
when an upgrade file is available for a new measurement cycle script.
Update the firmware
Use an SD card with an upgrade file to update the firmware for the controller, sensor or network card.
The upgrade menu is shown only when the SD card contains an upgrade file.
1.
Install the SD card into the SD card slot.
2.
Select SD CARD SETUP from the MAIN MENU.
Note: The SD CARD SETUP option shows only when an SD card is installed.
3.
Select UPGRADE SOFTWARE and confirm. Select the device and upgrade version, if applicable.
4.
When the upgrade is complete, the display shows TRANSFER COMPLETE. Remove the SD
card.
5.
Restart the instrument for the upgrade to take effect.
Calibration
N O T I C E
The manufacturer recommends calibration of the analyzer after 1 day of operation to allow all system components
to stabilize.
The automatic calibration uses the installed known standard to calibrate the analyzer. Calibration can
be done manually or scheduled to occur automatically. Use the calibration menu to view calibration
data, start or cancel the auto calibration, override the auto cal settings, calibrate manually or to reset
to the default calibration.
1.
Push
cal
to display the calibration menu.
Option
Description
START MANUAL CAL
Refer to
on page 18.
SET AUTO CALIBRATION
Refer to
Schedule automatic calibrations
on page 18.
English
17
Summary of Contents for 5500sc PO4 3-HR
Page 208: ...208 216 211 221 213 SD 224 213 224 214 1 1 1 6 2 7 3 8 4 9 5 Enter 2 3 208...
Page 209: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 4 0 100 1 209...
Page 210: ...45 75 10 45 50 75 10 50 2 6 1 2 3 ppb 48 SD 4 SD 210...
Page 211: ...4 SD 1 3 2 2 4 SD 1 1 1 2 3 4 211 4 1 2 3 4 2 211...
Page 212: ...5 6 7 0 14 bar 2 psi 0 28 bar 4 psi 55 mL 8 ENTER 9 10 224 216 216 1 enter enter 1 224 212...
Page 213: ...MSDS SDS 1 2 1 100 1 100 1 100 100 100 100 1 250 500 mL 2 213...
Page 214: ...3 4 5 5 5 1 2 3 4 211 1 2 3 9 10 240 15 ppb ppm mg L g L 1 5 1 16 10 214...
Page 215: ...10 216 223 216 4 20 mA 216 Modbus Profibus HART HACH55 215...
Page 218: ...PID 5 50 50 6 1 2 3 219 216 218...
Page 219: ...1 2 3 1 0 0 9 1 0 1 1 1 0 0 5 1 0 1 5 5 4 0 0 5 3 5 4 0 5 0 300 5 0 300 5 4 10 219...
Page 220: ...5 5 5 0 0 0 0 6 5 30 7 8 1 220...
Page 221: ...2 3 4 5 6 1 1 2 18 000 18 000 1 2 2 3 221...
Page 224: ...SD SD SD SD 2 GB 1 SD 4 211 2 SD SD SD 3 SD SD SD SD 1 SD 2 SD SD SD 3 4 SD 5 1 224...
Page 225: ...1 225 225 4 20 mA 1 2 3 ppb 4 1 2 ENTER 3 4 225...
Page 279: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 PROGNOSYS 4 PROGNOSYS PROGNOSYS 0 100 1 279...
Page 280: ...45 75 10 45 50 75 10 50 2 1 2 home 3 48 SD 4 SD 280...
Page 281: ...4 SD 1 3 2 2 4 SD 1 1 1 2 3 4 281 4 1 2 3 4 5 6 2 281...
Page 282: ...7 0 14 2 0 28 4 55 8 enter 9 10 294 286 286 1 enter enter 1 294 MSDS SDS 282...
Page 283: ...1 menu 2 1 100 1 100 1 100 100 100 100 1 250 500 2 3 4 5 5 5 283...
Page 284: ...1 2 3 4 281 1 2 menu 3 9 10 240 15 ppb ppm 1 5 1 16 10 10 284...
Page 285: ...285 292 286 4 20 286 Modbus Profibus HART HACH55 1 menu 2 285...
Page 287: ...4 20 1 menu 4 20 2 3 4 5 50 50 6 287...
Page 288: ...1 menu 2 3 288 286 1 menu 2 3 1 0 0 9 1 0 1 1 1 0 0 5 1 0 1 5 5 288...
Page 289: ...4 0 0 5 3 5 4 0 5 0 300 5 0 300 5 4 10 5 5 5 ONMAX 0 OFFMAX 0 ONMIN 0 min OFFMIN 0 min 6 289...
Page 290: ...5 30 7 8 1 2 3 4 5 6 1 1 menu 2 290...
Page 292: ...2 0 1023 511 3 A D REF A D A D 6 6 6 1 2 3 1 2 292...
Page 294: ...1 SD 2 SD SD SD 3 4 SD 5 1 1 cal 295 294 4 20 1 cal 2 3 ppb 4 294...
Page 295: ...1 cal 2 enter 3 4 295...
Page 370: ...4 SD 1 3 LED 2 2 4 SD 1 1 1 2 3 4 370 4 1 LANGUAGE 2 DATE FORMAT 3 DATE TIME 2 370...
Page 386: ...1 cal START MANUAL CAL 2 enter 3 YES NO 4 386...
Page 404: ......
Page 405: ......