Hach 2100Qis User Manual Download Page 1

DOC022.53.80041

2100Q and 2100Qis

12/2017, Edition 4

User Manual

Summary of Contents for 2100Qis

Page 1: ...DOC022 53 80041 2100Q and 2100Qis 12 2017 Edition 4 User Manual...

Page 2: ......

Page 3: ...face and navigation 7 User interface 7 Display description 8 Navigation 8 Startup 8 Turn the meter on and off 8 Change the language 8 Change the date and time 9 Standard operation 9 Use a sample ID 9...

Page 4: ...ion standard overview 14 StablCal RapidCal calibration 15 Verification options 16 Calibration verification Verify Cal 16 Reading modes 17 Apply silicone oil to a sample cell 17 Indexing a single cell...

Page 5: ...Data logger 500 records Specification Details Power requirement AC 100 240 V 50 60 Hz with power or USB power module 4 AA alkaline batteries Rechargeable NiMH for use with USB power module Operating c...

Page 6: ...tached to the instrument Personal injury or damage to the instrument could occur if not observed A symbol on the instrument is referenced in the manual with a precautionary statement This is the safet...

Page 7: ...at their expense The following techniques can be used to reduce interference problems 1 Disconnect the equipment from its power source to verify that it is or is not the source of the interference 2 I...

Page 8: ...artment Check the battery contacts for corrosion and clean them if necessary N O T I C E When using nickel metal hydride NiMH batteries the battery icon will not indicate a full charge after freshly c...

Page 9: ...BRATION key shows calibration screen start calibration select cal options 6 RIGHT key contextual read turbidity sample selects or confirms options opens jumps to sub menus 3 DOWN key scroll through me...

Page 10: ...o select an option There are two ways to change options 1 Select an option from a list Use the UP and DOWN keys to select an option If check boxes are shown more than one option can be selected Push t...

Page 11: ...s menu 2 Select create or delete a sample ID Option Description Current ID Select an ID from a list The current ID will be associated with sample data until a different ID is selected Create a New Sam...

Page 12: ...ode Signal Average mode or in the Rapidly Settling Turbidity mode Refer to Reading modes on page 17 for more information For accurate turbidity readings use clean sample cells and remove air bubbles d...

Page 13: ...ace Note Do not hold the instrument while making measurements 5 Gently invert and then insert the sample cell in the instrument cell compartment so the diamond or orientation mark aligns with the rais...

Page 14: ...entire Reading Log can be deleted at once Send stored data Data can be stored and transferred to a printer computer or USB storage device The data will be formatted as an XML file Install the USB powe...

Page 15: ...d 3 Select Security On to enable the password setting The requirement for the password entry is controlled by setting Security Options on or off Note Set the Security to Off to disable the password se...

Page 16: ...er to Table 1 for the calibration standard overview Table 1 Calibration overview Type of calibration Required standards StablCal RapidCal 0 40 NTU 20 NTU StablCal 0 1000 NTU 20 NTU 100 NTU 800 NTU For...

Page 17: ...urve 3 Select StablCal RapidCal from the list and push OK 4 Insert the 20 NTU StablCal Standard and close the lid Note The standard to be inserted is bordered 5 Push Read The display shows Stabilizing...

Page 18: ...one of the following options for time interval and push OK Off 30 min Default setting 2 h 4 h 8 h 24 h Allow Defer Push Allow Defer and select Yes or No to postpone the verification due time Calibrati...

Page 19: ...ng Turbidity reads and calculates five readings while showing intermediate results Apply silicone oil to a sample cell Sample cells and caps must be extremely clean and free from significant scratches...

Page 20: ...ll compartment Close the lid 5 Push the UP and DOWN key to access the Reading Options and then select Indexing Sample Cell Note The instruments always stays in the last selected reading mode 6 Push Re...

Page 21: ...Cap and store the sample cells 3 Wipe the outside of the sample cells dry with the a soft cloth Replace the battery W A R N I N G Explosion hazard An expired battery can cause hydrogen gas buildup in...

Page 22: ...1 2 3 4 20 English...

Page 23: ...5 6 7 8 English 21...

Page 24: ...tion Solution Temperature too high Switch off instrument Temperature has exceeded the meter limits 60 C or 140 F Turn off the meter and let it cool down RST Average value Solids are settling too slowl...

Page 25: ...cement parts Description Quantity Item no StablCal ampule calibration kit 1 2971205 10 NTU verification standard 100 mL 2961701 Replacement parts continued Description Quantity Item no Silicone Oil 15...

Page 26: ...StablCal 0 5 NTU Standard 100 mL 2698042 StablCal calibration kit 100 mL 2971210 StablCal calibration kit 500 mL 2971200 Gelex secondary standards set 1 2464105 Deionized water 4 vials 27217 Filter 0...

Page 27: ...display 8 I Indexing a sample cell 17 K keypad 7 L language 8 M menu navigation 8 meter components 5 meter specifications 3 meters troubleshooting 22 N normal reading 17 O operator ID 9 orientation m...

Page 28: ...V verification 16 Index 26...