Orbisphere 510/511/512 Analyzers - Inputs/Outputs Menu
69 of 138
Operator Manual
ORBISPHERE
6.3.1 Relays Configuration
The three relays (per channel) can be triggered by several
standard events, or a combination of events (custom).
The relays output can be used to turn on a beacon, horn
or PLC (see
“Connections to Electronic Boards” on
Note:
Relays can be set to Normally Open [NO] or Normally
Closed [NC] by changing the jumper positions on the
measurement board (see
).
Select a standard event in the rolling list
If “Custom event” has been selected, it has to be
configured by touching the
configure
button
Press on the text box to open the selection menu
(rolling menu). Select the events that must trigger the
relay, and press OK.
The example shown here will trigger the relays whenever
the value is above the High High or under the Low Low
preset values.
Proceed in the same manner for any other event that
should trigger the relay
Summary of Contents for ORBISPHERE 510
Page 1: ...Operator Manual ORBISPHERE 510 511 512 Revision G 05 11 2008 ...
Page 2: ......
Page 12: ...6 of 138 Table of Contents Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 48: ...42 of 138 View Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 86: ...80 of 138 Inputs Outputs Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 110: ...104 of 138 Security Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 134: ...128 of 138 Specifications Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 140: ...134 of 138 Glossary Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 143: ...Orbisphere 510 511 512 Analyzers User Notes 137 of 138 Operator Manual ORBISPHERE User Notes ...
Page 144: ......