Other maintenance options
1.
Select an option, then push
Select
.
Option
Description
Stirring activation
Do a check of the magnetic stirrer. Push the up and down arrow keys to increase or
decrease the stirring speed.
Live measure
This option is not available for all sensors. The option shows continuous
measurement data with connected sensors to quickly check measurements. The
installed applications and the automatic additions to the sample are set to off.
Continuous measurements are not temperature compensated, so measurement
differences may occur in the same sample between continuous measurements and
measurements that use installed applications with temperature compensation.
Maintenance
summary
See the number of days remaining for maintenance tasks. After doing a task, push
Reset
to set the number of days remaining to the default value.
Maintenance
schedule
See the list of maintenance tasks. Push
Edit
to change the default value.
Reagent
replacement
Use this option to replace the reagents. Obey the instructions on the display.
Troubleshooting
Refer to the following table for common problem messages or symptoms and possible causes.
Error/Warning
Description
Stop requested
The operator has pushed the stop button
Measurement out of range
The measurement is out of range
Temperature out of range
The temperature measurement is out of range
Temperature unknown
The temperature measurement cannot be determined
Temperature variation too high
The variation in temperature is too high
Out of range
The computed result is outside accepted limits
Equation calc. error
An equation variable is unknown
Titrant calibration has expired
The Calibration date has expired
Maximum number of applications is reached
The maximum number of applications per line has been
reached
Requires at least one titrant that cannot be
installed
The application is not compatible with applications already
installed as it uses a different titration
Calibration solution already used
The calibration solution has already been used
No titrant information
System failure. Contact technical support
No measure received
System failure. Contact technical support
Burette fails to deliver
The syringe cannot be emptied
Burette fails to fill
The syringe cannot be filled
Burette failed to read delivered volume
System failure. Contact technical support
Excess of titrant
Too much titrant during cell conditioning
Excess of water
Too much water during cell conditioning
22
English
Summary of Contents for TitraLab KF1121
Page 21: ...English 21...
Page 130: ...130 Portugu s...
Page 151: ...e tina 151...
Page 172: ...172 Dansk...
Page 237: ...Svenska 237...
Page 262: ...1 2 15 1 2 3 4 5 1 262...
Page 263: ...1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 KF1121 1 2 2 USB USB 263...
Page 264: ...USB USB 2 1 24 V 4 7 USB 2 1 5 8 Ethernet 3 2 6 3 1 3 3x 5 2 4 264...
Page 265: ...4 1 KF 6 2 KF Teflon 7 3 10x 8 4 USB 9 3 x GL45 5 10 3x 1 15 6 in 1 2 1 265...
Page 266: ...GFCI GFI 1 2 2 264 3 180 266...
Page 267: ...1 263 267...
Page 268: ...268...
Page 269: ...269...
Page 270: ...MSDS SDS 270...
Page 271: ...278 1 2 271...
Page 272: ...5 5 1 4 2 5 3 2 272...
Page 273: ...MSDS SDS 1 Settings 2 Select 273 273 3 Back 1 Settings 2 Applications Edit 273...
Page 275: ...MSDS SDS a 1 2 3 USB 4 276 5 Start 6 7 8 Start 9 10 Skip Stop 11 12 13 275...
Page 277: ...1 Data log 2 Select USB 3 1 Calibration 2 3 Start 4 5 6 Continue 7 Yes No 8 Exit 277...
Page 278: ...274 300 C 572 F 4 Maintenance 1 Pump activation 278...
Page 279: ...Syringe replacement 266 1 Pump activation 2 Toggle 279...
Page 280: ...1 Select Reset 280...
Page 281: ...Edit 281...
Page 322: ...322 Rom n...
Page 343: ...T rk e 343...
Page 364: ...364 Slovensk jazyk...
Page 385: ...Slovenski 385...
Page 406: ...406 Hrvatski...
Page 410: ...1 2 15 FCC 1 2 3 4 5 1 410...
Page 411: ...1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 KF1121 1 2 2 USB USB 411...
Page 412: ...USB USB 2 1 24 V 4 7 USB 2 1 5 8 Ethernet 3 2 6 3 1 3 x3 5 2 4 412...
Page 413: ...4 1 KF 6 2 KF Teflon 7 3 x10 8 4 USB 9 3 x GL45 5 10 3x 1 15 cm 1 2 1 413...
Page 414: ...AC GFCI GFI 1 2 2 412 3 180 414...
Page 415: ...1 411 415...
Page 416: ...416...
Page 417: ...417...
Page 418: ...MSDS SDS 418...
Page 419: ...426 1 2 419...
Page 420: ...5 5 1 4 2 5 3 2 420...
Page 423: ...MSDS SDS 1 2 3 USB 4 424 5 Start 6 7 8 Start 9 10 Skip Stop 11 423...
Page 426: ...3 Start 4 5 6 Continue 7 Yes No 8 Exit 422 300 C 4 426...
Page 428: ...428...
Page 431: ......