14
Installation
3.7 Flow chamber installation
The model 32013A or 32015 flow chamber allows the liquid or gaseous sample past the TC and
temperature sensors. The flow chamber attaches to the TC sensor with a threaded collar and is
sealed to the sensor with two O-rings (EPDM
∅
20x2 mm and NBR
∅
4x1.5 mm).
CAUTION
At the top of the TC sensor is a thermistor to measure the sample temperature. Do not
bend or damage this component.
The flow chamber’s inlet and outlet use ¼-inch diameter transparent plastic tubing. Connect by
compression fittings to the sample source and to the drain, respectively.
You may also have received a model 32051 sample tube adapter to attach the flow chamber
inlet tubing to the sampling point (see
WIN3654 PC program installation on page 11
).
Figure 4 Flow chamber 32013A