
50
Maintenance and Troubleshooting
8.6.4 Troubleshooting table
Table 6 Troubleshooting table
Symptom
Cause
Possible Solution
Unstable
measurement
Degassing
Adjust sample flow rate
Flow chamber and/or membrane not clean
Clean system
Sensor’s membrane not tight smooth and
wrinkle free
Replace membrane and
recalibrate
Leaking solenoid valve
No gas bubbles should escape from
the purge gas outlet during
measurement. If the solenoid leaks
(more than 1 bubble in 15 seconds)
contact your Hach Lange
representative.
An external power supply is used but the
power supply is not clean
Switch to battery power
Low readings
Sample flow rate too low
Check flow rate
Degassing
Adjust forcing gas and/or sample flow
rate
Membrane is not at sample temperature
Allow sample to flow past the
membrane to make sure the
membrane is at sample temperature
before making measurements
Membrane not clean
Clean system
Incorrect solubility curve
Change solubility
Incorrect calibration
Recalibrate
High readings
Membrane is not at sample temperature
Allow sample to flow past the
membrane to make sure the
membrane is at sample temperature
before making measurements
Incorrect solubility curve
Change solubility
Incorrect calibration
Recalibrate
Summary of Contents for ORBISPHERE 3654
Page 1: ...ORBISPHERE Model 3654 Portable Analyzer USER MANUAL June 2009 Revision I ...
Page 2: ......
Page 10: ...8 General Information ...
Page 18: ...16 Installation ...
Page 28: ...26 Operating Instructions ...
Page 40: ...38 Calibrations ...
Page 42: ...40 Accessories and Attachments ...
Page 56: ...54 Glossary of Terms ...
Page 57: ......
Page 58: ......