10
Specifications
2.2 Sensor specifications
Table 1 Sensor specifications
Sensor Type
31 270 H
2
Membranes 29561A
2952A
2935A
Thickness
25 µm
25 µm
25 µm
Membrane material
PFA
ETFE
ECTFE (Halar)
Recommended applications
Waste gas offgas,
reactor coolant
Reactor coolant
High H
2
level
Radiation limits
10
5
rad
10
8
rad
10
8
rad
Measurement range at 25°C
0-1.5 ppm,
or 0-16 cc/kg,
or 0-1 bar
0-5 ppm,
or 0-60 cc/kg,
or 0-3 bar
0-15 ppm,
or 0-180 cc/kg,
or 0-9 bar
Accuracy (sample temp.
20-50°C within ± 5°C of
calibration temperature)
The greater of
±1% of reading
or ± 4 ppb,
or ± 0.06cc/kg,
or ± 3 mbar
The greater of
±1% of reading
or ± 16 ppb,
or ± 0.2cc/kg,
or ± 12 mbar
The greater of
±1% of reading
or ± 45 ppb,
or ± 0.5 cc/kg,
or ± 30 mbar
Accuracy (sample temp.
0-50°C independent of
calibration temperature)
The greater of
±2% of reading
or ± 30 ppb
or ± 0.35 cc/kg
or ± 12 mbar
The greater of
±2% of reading
or ± 120 ppb
or ± 1.2 cc/kg,
or ± 40 mbar
The greater of
±2% of reading
or ± 300 ppb
or ± 3.0 cc/kg,
or ± 100 mbar
Measurement cycle (sec.)
20
Sample flow rate (through
flow chamber)
200 ml/min
100 ml/min
50 ml/min
Table 2 Sensor Specifications
Sensor Type
31 272 H
2
31 570 N
2
Membranes 29561A
2952A
2935A
29561A
Thickness
25 µm
25 µm
25 µm
25 µm
Membrane material
PFA
ETFE
ECTFE (Halar)
PFA
Recommended
applications
Waste gas and
offgas
Waste gas offgas,
reactor coolant
Reactor coolant
and High H
2
level
Beverage
Radiation limits
10
5
rad
10
8
rad
10
8
rad
10
5
rad
Measurement range at 25°C
0-0.075 ppm,
or 0-8 cc/kg,
or 0-0.5 bar
0-2.5 ppm,
or 0-30 cc/kg,
or 0-1.5 bar
0-7.5 ppm,
or 0-90 cc/kg,
or 0-4.5 bar
0-250 ppm,
or 0-200 ml/L,
or 0-15 bar
Accuracy (sample temp.
20-50°C within ± 5°C of
calibration temperature)
The greater of
±1% of reading
or ± 2 ppb,
or ± 0.03 cc/kg,
or ± 2 mbar
The greater of
±1% of reading
or ± 8 ppb,
or ± 0.1cc/kg,
or ± 6 mbar
The greater of
±1% of reading
or ± 25 ppb,
or ± 0.25 cc/kg,
or ± 15 mbar
The greater of
±2% of reading
or ± 0.5 ppm,
or ± 0.5 ml/L,
or ± 30 mbar
Accuracy (sample temp.
0-50°C independent of
calibration temperature)
The greater of
±2% of reading
or ± 60 ppb
or ± 0.7 cc/kg
or ± 25 mbar
The greater of
±2% of reading
or ± 250 ppb
or ± 2.5 cc/kg,
or ± 80 mbar
The greater of
±2% of reading
or ± 600 ppb
or ± 6.0 cc/kg,
or ± 200 mbar
The greater of
±4% of reading
or ± 2 ppm,
or ± 1.6 ml/L,
or ± 70 mbar
Measurement cycle (sec.)
20
Sample flow rate (through
flow chamber)
200 ml/min
100 ml/min
50 ml/min
150 ml/min
Summary of Contents for ORBISPHERE 3654
Page 1: ...ORBISPHERE Model 3654 Portable Analyzer USER MANUAL June 2009 Revision I ...
Page 2: ......
Page 10: ...8 General Information ...
Page 18: ...16 Installation ...
Page 28: ...26 Operating Instructions ...
Page 40: ...38 Calibrations ...
Page 42: ...40 Accessories and Attachments ...
Page 56: ...54 Glossary of Terms ...
Page 57: ......
Page 58: ......