Guide to the troubleshooting – USC
Symptom
Current output
Event
code
Cause
Resolution
None
Not defined
IU00
Switch-on
Registration of the switch-on time
IU01
Additional module installed
Registration of the installed
additional module
IU02
Additional module defective or removed
Replace or install additional module
IU03
Parameter successfully corrected
Information
No display
1
Supply voltage missing/too low
2
USC 5000/7000 defective
1
Check supply voltage
2
Replace USC 5000/7000
Dots in the uppermost
line of the display
(I OUT 1)
EU61
Refer to
(I OUT 2)
EU62
Refer to
(I OUT 3)
EU63
Refer to
(I OUT 4)
EU64
Refer to