English
115
Guide to the troubleshooting - single
position transmitter
OXY 4100/3150/4150
Symptom
Current signal Error
Resolution
Constant electric signal
(varying oxygen
concentration)
I = 0 mA
Incorrect wiring
Check wiring, and remedy if necessary
Voltage too low
Check supply voltage at the transmitter.
Supply voltage must be at least 12 V
Initialization error
1
Switch off transmitter for at least 5 seconds
2
Switch on again with sensor pointing downward
Transmitter defective
Replace transmitter
I = 3.80 mA
or 24.0 mA
Initialization error
1
Switch off transmitter for at least 5 seconds
2
Switch on again with sensor pointing downward
Transmitter defective
Replace transmitter
I = 3.85 mA
or 23.5 mA
Membrane damaged
Replace OXY 1100 sensor
Moisture behind OXY 1100 sensor
Fully dry out the terminals on the transmitter. A
hair dryer can be used
I = 3.90 mA
or 23.0 mA
Calibration not possible, worn out OXY 1100 sensor
Replace OXY 1100 sensor
I = 3.95 mA
Oxygen content below set measurement range
Reduce lower measurement limit value
I = 22.00 mA
Oxygen content below set measurement range
Raise upper measurement limit value
Incorrect measurement
value
Variable
Incorrect calibration value
1
Check full-scale value in atmospheric air. Hold the
transmitter in the open air with the sensor downward.
Compare the value with DS/EN 25814 (refer to
appendix IIa)
2
Check the zero point (refer to appendix IIb)
3
If significant deviation, calibrate again.