| SERVICE MANUAL / INSTALLATION GUIDE
8
© HAAG-STREIT AG, 3098 Koeniz, Switzerland
–
HS-Doc. No. 7220676 - Edition 2.00 / 2020
3.2
Charging the FUNDUS MODULE 300
WARNING!
•
Place cradle outside the patient environment (at least 1.5 meters distance from the patient).
•
Use only medically approved PCs or operate via medically approved isolating transformer.
•
If the PC connected to a customer network make sure to apply appropriate safety measures,
such as Antivirus and Firewall protection.
Place the cradle on a desk next to the PC. The cradle
shall be outside the patient environment at least 1.5
meters distance from the patient.
Fig. 7
Connect the power supply cable to the cradle.
Connect the cradle USB cable between the PC and
the cradle. Avoid using excessive force or twisting the
connector when connecting the USB cord to a PC.
Fig. 8
Guide the cables underneath the cradle through the
cable holders and plug the power supply cable to the
mains.
Fig. 9