8 | Troubleshooting
Haag-Streit AG
34
1500 4000089 02010_web
8 Troubleshooting
CAUTION
Risk of limited treatment options!
If defects are not detected, the surgical microscope system may not offer all functions during sur-
gery.
a) Before each use, make sure that all electrical and mechanical connections are firmly seated
and free of defects.
b) Regularly check the functionality and quality of the magnified vision.
c) Do not use the system in case of any obvious defects, error states or functional restrictions.
Call the Haag-Streit Service.
NOTICE
Resettable error reoccurs twice or more
The error may be due to a systematic fault in the medical device and must be attended by
Haag-Streit.
Please contact your local representative or Haag-Streit, stating the product name, type and serial
number together with a description of the error and the circumstances under which the error oc-
curs.
Error messages
Cause
Remedy
Error message
HEAD ER-
ROR
appears.
The CCU does not de-
tect the camera head.
Check the connection of the camera cable
to the camera head and to the CCU.
Error message
AWB ERROR
appears.
Automatic white bal-
ance was not per-
formed because the il-
lumination is too bright
or too dark.
Adjust the illumination and perform the
automatic white balance again.