safety, installation and commissioning
file: n:\article\cos30022man-en\20140206_cos30022man-en_h-p-cosmos_robowalk_expander_manual.doc
© 2014 h/p/cosmos sports & medical gmbh author: fh [email protected] created 06.02.2014 printed 06.02.2014 page: 22 of 67
8
Installation
8.1.1
Unpacking and packaging
When receiving the device in a crate or unpacked, make sure the device, the accessories and/or the packaging is not
damaged. If you discover any damage and/or missing parts make a note on the packing-list / delivery note of the carrier.
Inform h/p/cosmos and your dealer immediately in writing about any damage and/or missing parts.
The manufacturer does not assume liability for any damage, complaints or missing parts which are
not reported immediately upon delivery on the packing list/delivery note.
Before unpacking the machine and accessories, read the instructions on the crate. Pay special attention to small parts,
so that you do not dispose of them or any instructions with the packaging. Within Germany, most devices are delivered
and assembled by h/p/cosmos directly or by an authorised forwarder. If delivered by h/p/cosmos, the packaging will be
removed and recycled.
If the device is delivered by a carrier, you can recycle the packaging yourself or send it back to the manufacturer
(transportation to be paid by the customer).
8.1.2
Transport
The h/p/cosmos robowalk
®
expander system is usually delivered in two h/p/cosmos standard boxes and with an
h/p/cosmos treadmill (see shipment instruction in the treadmill manual).
Some of the components are heavy in weight!
Please notice that the installation must be done by h/p/cosmos authorized technicians.
8.1.3
Storage
Store the devices at a temperature of – 20° ... + 50° C.
8.1.4
Environmental requirements / room configuration
Running machines are not to be used in medically utilized rooms with a danger of explosions or in easily inflammable
atmospheres. The devices must not be installed near to e.g. an x-ray device, motors or transformers with high voltage
connection, as the electric and magnetic interference can falsify measurements or even render them impossible. High
voltage lines must be avoided nearby the device. h/p/cosmos electrical devices with mains connections must neither be
used in wet and humid areas (e.g. swimming pools, saunas, etc.) nor in environmental chambers. If not stated otherwise
in the delivery information, h/p/cosmos devices are designed for operation in normal climatic surroundings:
Temperature:
+ 10° ... + 40°C
Relative humidity:
30 ... 70 % (non condensing!)
Air pressure:
700 ... 1060 mbar
Maximum operating altitude: approx. 10,000 feet (3000 m), without pressurization
The device should be protected from high humidity.