Maintenance and safety inspections
N
Maintenance and repair of the devices are to be performed only by service engineers authorized
and certified by h/p/cosmos, preferably within the scope of a maintenance contract.
N
In case of a detected and/or assumed malfunction and/or defects and/or unreadable safety labels,
the device is to be disengaged immediately. Mark the device and ensure that it cannot be
operated. Inform the supplier and authorized service personnel in writing immediately.
N
Disregarding warnings, notifications of intended and forbidden use, precautions as well as
unauthorized or lack of maintenance and/or regular safety checks may lead to injuries or death
and/or can damage the device. Furthermore it will result in loss of any liability and warranty.
N
During all maintenance work and safety tests make sure that no third parties are directly or
indirectly in contact with the device under test and/or the technician performing the test. Keep a
safety zone of 2 m radius clear.
9.1
Preventive maintenance
The h/p/cosmos authorised service engineers are happy to help you in case problems occur.
Preventive maintenance can prevent future problems and is indispensable for the safety of technical devices. Therefore, ask
our service department for an annual preventive maintenance contract, which is highly recommended by the manufacturer.
Some basic regular maintenance and regular safety checks as stipulated in the following chapters are obligatory!
Before starting to work with the device, the user always has to check visually:
N
expanders
N
hooks
N
arretierelemente
N
labelling
N
leg cuffs
9.2
Immediate maintenance
Immediate maintenance is necessary:
N
if a defect or malfunction of the device has been detected or is suspected at preventive maintenance
N
every 12 months.
Only a properly and regularly serviced device is safe.