page 24 of 99
cos105000_170-190_65, rev. 1.03
© 1988 - 2019 h/p/cosmos sports & medical gmbh
4.4 Emergency off
Do only use in case of emergency.
Do not use as normal stop button.
Do not stand on or enter the running deck when device is in elevation (running belt might slip through due to gravity).
The operator has to be in reach of the emergency off at any time.
If the operator is not able to reach the emergency off button at the UserTerminal (body height, obstacles, etc.), the operator must
install an additional emergency stop within reach (see accessories).
4.5 Emergency stop
Do only use in case of emergency.
Do not use as normal stop button.
Do not stand on or enter the running deck when device is in elevation (running belt might slip through due to gravity).
Operation
Result
Release
Restart
UserT
erminal
Power supply to load
generating components
interrupted
Running belt spins out (until
0 km/h) with
undefined deceleration
Movement of elevation
system stops
UserTerminal off
Interface communication
interrupted
Switch the device on.
Push button
Pull button
Restart application.
Operation
Result
Release
Restart
Safety arch
Running belt stops with
predefined deceleration
Movement of elevation
system stops
UserTerminal displays
“pull stop”
Mains connection and
Interface communication
not interrupted
Restart application
Pull cord
Release cord
Pull cord safety stop
Pull cord with magnet from
metal plate
Reconnect magnet with
cord to metal plate