Version 252 of 2018-04-06
!
3.3
Functional checks
Set the pressure to 2.0 bar
Reset the actual number of pulses on the parameter display of the operating
panel.
Release pulses with a frequency of 5Hz
Check that the triggered shocks are correctly counted on the treatment shock
counter of the ShockMaster control unit.
3.4
Standard operation
Connect the V-ACTOR II handpiece to the control unit
Set the pressure to an initial value of 2.0 bar and a frequency of 20Hz.
Start the V-ACTOR II treatment by pushing the trigger button.
3.5
Treatment
CAUTION
Always read chapter 3.1 SAFETY INSTRUCTIONS before starting a treatment.
NOTE
The use of maximal pressure level during treatment must not result in any uncomfortable pain
for the patient.
Treatment may be applied directly to the patient’s skin or with textile as
intermedium.
Perform V-ACTOR II treatment as described in the application recommendations.