OPERATION
31
Sense Connection
1.
Remote sense can be used to help compensate
for long cable length. The longer the cable, the
higher the potential resistance and inductance,
therefore a short cable is best. Twisting the
cable can help reduce induced inductance and
using the Vsense terminals compensates the
voltage drop seen across the load leads,
especially leads with higher resistance. This is
useful when used in CV, CR or CP mode.
Make sure the master unit and booster packs
are turned off.
1.
Make sure the DUT is turned off.
2.
Connect the sense cables from the DUT to the
master unit
Connect a sense cable from the positive sense
terminal to the positive potential of the DUT.
Connect a sense cable from the negative sense
terminal to negative potential on the DUT.
Do not connect the sense cables to the booster
packs.