PEL-3211(H) User Manual
28
Connection
The PEL-3111(H) can control up to 4 PEL-3211(H) booster units in
parallel. Excluding the master unit, other PEL-3000(H) units cannot
be used in parallel with the booster packs.
Master unit settings for response rate, response speed and so on
will be adopted by any connected booster packs.
Frame Link Connections
Description
The J1 and J2 connectors on the rear panel are
used to control the booster packs for parallel
operation. The GTL-255 cable is used to connect
the J1 and J2 connectors. The GTL-255 frame link
cable is used for units that are stacked vertically.
Note
Ensure that the frame link cables are separated as
much as possible from the load cables to prevent
unstable operation.
Connection
1.
Make sure the master unit and booster packs
are turned off.
2.
Make sure the DUT is turned off.
3.
Connect the Master unit to the booster packs
via the J1/J2 connectors*.
Use the GTL-255 frame link cables.
Connect from:
Master J2
Slave1 J1
Slave1 J2
Slave2 J1 and so on.
Remove one ferrite core from the last frame link
cable. Remove the ferrite core that is closest to
the J1 port on the last slave unit. See the
diagram below for details.