PEL-3211(H) Booster Pack Introduction
The PEL-3211(H) Booster is designed to be used with the PEL-
3111(H) Electronic Load for parallel load applications to increase
the current sinking capacity. Up to 4 booster packs can be
connected as slave units to a PEL-3111(H), which acts as the master
unit.
Please note that throughout this manual the term “PEL-3000(H)”
refers to any one of the models in the series lineup, unless
specifically stated otherwise. The termer “booster” refers only to
the PEL-3211(H). The term “master” or “master unit” refers to the
PEL-3111(H) unit that is configured as the master unit when in
parallel mode.
Booster Capacity
The table below shows the total current and power capacity by the
number of booster packs connected to a PEL-3111(H) master unit.
No. of
Boosters
Total Current:
BMaster
(PEL-3211(H)+PEL-3111(H))
Total Power:
BMaster
(PEL-3211(H)+PEL-3111(H))
1
630A (420A+210A)
(H)157.5A(105A+52.5A)
3150W (2100W+1050W)
2
1050A (840A+210A)
(H)262.5A(210A+52.5A)
5250W (4200W+1050W)
3
1470A (1260A+210A)
(H)367.5A(315A+52.5A)
7350W (6300W+1050W)
4
1890A (1680A+210A)
(H)472.5A(420A+52.5A)
9450W (8400W+1050W)