OPERATION
179
Configure
Previous
Menu
Protection
Go-NoGo
CH1
Other
Von Voltage
Von Latch
CH CONT
Independent
CC Vrange
V
0.00
OFF
Panel
OFF
High
conf
Group
CCDH
80V
LOAD
USB
Load D-Time
0.0 S
05/04/22
16 : 50
2.
Use the Selector knob to highlight
CC Vrange.
3.
Press the Selector knob to edit CC
Vrange, then turn to increase or
decrease the range.
CC Vrange
High
4.
Press the Selector knob or Enter
to confirm selection.
ENTER
or
Note
The configuration settings only apply to the current
channel.
Adjusting the Von Voltage and Latch
Background
The Von Voltage is the voltage point at which the
load module will start to sink current. When Von
latch is set to ON, the load will continue to sink
current after being tripped, even if the voltage
drops below the Von Voltage level. The step
resolution of Von Voltage is load module
dependent.