OPERATION
145
Select CP Range
Background
Constant Power mode can run in high and low
range. The maximum range is dependent on the
load module. Some models are only high range.
Ensure the menu is in CP Mode. See page 138.
Configure
Configure
Mode
CP
Range
Low
Panel Operation
1.
Press the F2 (Range) key repeatedly
until High or Low range is selected.
F1
F5
~
F2
Range
High
Range
Low
The range will be reflected in both the
bottom menu system and the Current
Operation Channel Status panel.
CPL
CH1
main
80V
CP High Range
CV Low Range
CPH
CPL
Note
Changing the range will only affect the current
(active) channel. Other channels will not be affected
by any changes.
Not all load modules support dual ranges. If only one
range is supported, it is usually high range.