COMMAND SYNTAX
298
<n> = 0 to 210 Expected reading in volts
DEFault 21V
MINimum 210mV
MAXimum 210V
UP Select next higher
measurement range
DOWN Select next lower
measurement range
Example
:SENSe1:VOLTage:DC:RANGe:UPPer DEFault
Command
[:SENSe[1]]:VOLTage[:DC]:RANGe?
Function
:RANGe? Query measurement range.
:RANGe? DEFault Query *RST default range.
:RANGe? MINimum Query lowest range (returns
0).
:RANGe? MAXimum Query highest range.
Example
:SENSe1:VOLTage:DC:RANGe?
Command [:SENSe[1]]:RESistance:RANGe[:UPPer]
<n>|UP|DOWN
Function
This Command is used to manually select the
measurement range for the specified measurement
function. The range is selected by specifying the
expected reading. The instrument will then go to the
most sensitive reading that will accommodate that
reading. For example, if you expect a reading of
approximately 50mV, then simply let <n>= 0.05 (or
50e-3) in order to select the 200mV range. You can
also use the UP and DOWN parameters to select
range. Each time UP or DOWN is sent, the next