Command
:SOURce[1]:LIST:CURRent:POINts?
Function
Query length of I-Source lis
Example
:SOURce:LIST:CURRent:POINts?
Command
:SOURce[1]:LIST:VOLTage:POINts?
Function
Query length of V-Source list
Example
:SOURce:LIST:VOLTage:POINts?
Command
:SOURce[1]:LIST:CURRent:STARt <n>
Function
This Command sets the starting point in the current
list sweep. The <n> parameter is 1-based, and it must
be less than or equal to the number of points in the
sweep. The new starting location will only be used
when the direction of the sweep is up
(:SOUR:LIST:CURR:DIR UP, for example). Changing
the direction to down will make the sweep start at
the last point in the list, but the starting point will be
restored when the direction is changed to up. If the
sweep reaches the end of the list, the sweep will
continue from the first point in the list. The starting
points for voltage and current sweeps are saved
separately, and are part of a user-saved setup.
<n> = 1 to 100 Set start point of list sweep
MINimum 1
MAXimum 100
DEFault 1
Example
:SOURce:LIST:CURRent:STARt 1
Command
:SOURce[1]:LIST:VOLTage:STARt <n>