GDS-200 & GDS-300 Series User Manual
176
Fuse Replacement
Background
The GDS-200/300 has fuse protection for both
current ports, a 10A and a 1A fuse.
Fuse Type
1A: T1A 500V
10A: T10A 500V
Steps
1.
Turn the power switch off on the GDS-200/300.
2.
Remove the 4 screws that hold the back panel
in place.
3.
Remove the back panel carefully. You will need
a case opening tool to pry open the case after
the screws are removed.
Note: the fan wiring is connected to the back
panel.
4.
Replace the fuse that has been blown, as shown
below (bottom right-hand side).
1A fuse
10A fuse