AQUASCAN 610
13/41
Version 1.1
Fig. 11 Resultant filter
The correlation curve and the result will be updated while changing the
filter.
4.4.3.
Notch filter:
Switch notch filter on or off. When selecting
on
you can enter the notch
filter frequency. Notch filter can be used for suppressing electrical
mains disturbing signals that belong to a small frequency range, e.g.
nearby A/C currents at 50/60 Hz. As a rule 110/120V AC creates a
50Hz disturbance and 230/240V AC create a 60Hz disturbance.
4.4.4.
Multiple filters:
Fig. 12 Multiple filter