Section 4
TM9230A-C1-002 November, 2021
2 November, 2021
4-5
I
current
= √(P
ower
/R
esistance
)
Currents for Natural Convection
(A)
Currents for 92310 Forced Convection
(A)
Model
Value (Ω)
5 W
10 W
Max A
10 W
25 W
Max A
9230A-10
1.0
2.2
3.2
5
3.2
5
≤10
1
9230A-15
0.5
3.2
4.5
7.5
4.5
7.1
15
9230A-15R
0.1
7.1
10.0
15
Rated only 25 W
9230A-30
0.1
7.1
10.0
15
10.0
15.8
30
9230A-50
0.05
10.0
14.1
25
14.1
22.4
50
9230A-100
0.01
22.4
31.6
50
31.6
50
100
9230A-150
0.005
32
45
75
45
71
150
9230A-300
0.001
71
100
150
100
158
300
9230A-500
0.5 m
100
141
250
141
224
500
9230A-1000
0.1 m
224
316
500
316
500
1000
9230A-1500
0.05 m
316
447
750
447
707
1500
9230A-3000
10
µ
707
1000
1500
1000
1581
3000
9230A-5000
5
µ
1000
1414
2500
1414
2236
5000
Table 4-1 : Test Points and Currents