MAINTENANCE
23
GMG-200
310-2310-001 EN20
Display
Description
Remedy
Blank display - device cannot
be switched on
Check power supply or
battery charge;
charge battery if necessary
Battery empty !
Start: pull in
Self-discharge protection:
The device has an automatic
shutdown feature to prevent
deep discharge
Completely retract the strap.
Out of memory
read or erase !
The memory is so full that no
more complete measuring
series can be saved
Press the “Clear memory” button for five
seconds
12 Maintenance
CAUTION!
If overload occurs during use of the device, e.g. due to blockage during the mea-
suring run, the measurement mechanism is subjected to extremely high forces.
−
It is advisable to have the device inspected at the plant!
12.1 Cleaning
The device is designed such that moisture that penetrates the device via the traction cable
does not impair operation. It is advisable to keep the traction cable tensed during measure-
ments to prevent excessive moisture from penetrating the device.
When all measurements are finished, extract the traction cable all the way, wipe it dry with a
soft cloth and then retract it again.
Caution!
The steel strap on the GMG-200 may have sharp edges, thus posing a risk of
cuts.
−
Wear gloves when drying!