Tympanometry
Pressing the Tymp button displays the tympanometry screen. The tympanometry
screen may also be selected from the Title Bar test type drop down in the center
of the title bar. The diagnostic tympanometry screen is show below.
The tympanometry screen presents information for conducting a diagnostic
tympanogram. Individual tympanogram results are stored as tabs on the left side
of the tympanogram graph. A check mark indicates the ‘marked’ tympanogram.
The marked tympanogram is the tympanogram that will be transferred to GSI
Suite or printed directly from the device. The peak pressure of the marked
tympanogram is used to set the pressure for conducting the acoustic reflex tests.
Only a single tympanogram per probe tone frequency may be marked. A total of
six tympanogram tabs per ear can be displayed on the screen. The tabs at the top
of the tympanogram graph are used to select the right or left ear. The current
selection is indicated by the larger tab.
Up to three tympanograms may be displayed per test. When running multiple
tympanograms the direction of the pressure sweep must alternate between sweeps
(i.e. if the first sweep goes from positive pressure to negative pressure the second