Chapter 6: Configuring a Temperature Zone
Calibrating the Temperature Probe
Although temperature probes are manufactured with high accuracy, some probes might show slight meas-
urement variations. The probe calibration allows you to compensate for such a variation by up to ± 10 °
F (± 5.5 °C) with an accuracy of one tenth of a degree
1. In edit mode, under the
Calibration
tab, click on the edit field next to the
Offset
button to display the
keypad.
2. Enter the impedance offset in ohms.
3. Enter the temperature offset and the low and high trouble impedances.
NOTE:
Enter value from left to right and always enter a decimal value. For example, a value of 1 °F
requires you to enter 10 to obtain 1.0 °F.
Use the
+/-
button to change between a positive and a negative value.
For example, if a probe shows a measurement difference of + 1.3 °F from the actual temperature, you
must enter – 1.3 °F to calibrate the probe.
38
895–00677
Agri-Alert 128 Touch
Summary of Contents for 890-00583
Page 6: ...NOTES 6 895 00677 Agri Alert 128 Touch...
Page 14: ...NOTES 14 895 00677 Agri Alert 128 Touch...
Page 26: ...NOTES 26 895 00677 Agri Alert 128 Touch...
Page 54: ...Chapter 8 Configuring a 0 5 Volt Zone Type 54 895 00677 Agri Alert 128 Touch...
Page 56: ...NOTES 56 895 00677 Agri Alert 128 Touch...
Page 64: ...NOTES 64 895 00677 Agri Alert 128 Touch...
Page 68: ...NOTES 68 895 00677 Agri Alert 128 Touch...
Page 72: ...NOTES 72 895 00677 Agri Alert 128 Touch...
Page 75: ...B Wiring Diagram with Terminals 895 00677 Agri Alert 128 Touch 75...
Page 78: ...NOTES 78 895 00677 Agri Alert 128 Touch...
Page 84: ......