14.3.2 Dismantling the measuring cell

1. Remove the housing cylinder and ascending pipe.

– Remove the housing cylinder (T), and dismantle it.
– Remove the ascending pipe (S). 

2. Remove the sample water regulating device and float body.

– Remove the valve insert (Q2) and adjusting spindle (Q1), 

and separate them.

– Remove the float body (N).

3. Remove the filter.

– Unscrew the filter cartridge (P).
– Slacken the screw cap for the filter (P1) at the upper end of 

the filter cartridge (P).

– Remove the filter strainer (P2) from the filter 

receptacle (P3).

4. Remove the calibration cup and evacuation spindle.

– Remove the calibration cup (I) and evacuation spindle (J), 

and separate them.

5. Remove the shut-off spindles and the bleeding spindle.

– Unscrew the shut-off spindles of the sample water inlet (V1) 

and outlet (V2) and the bleeding spindle (L).

14.3.3 Cleaning the measuring cell

• Required tools and aids:

– soap suds, possibly diluted hydrochloric acid (5-10 %), 

water

– abrasive domestic cleaning agent (for instance ATA, VIM).
– The flow armature (F) can be cleaned with conventional 

soap suds. 

– If high contamination or furring is evident, clean the 

dismantled flow armature (F) briefly with diluted hydrochloric 
acid (max. 5-10 %).

1. Clean the metal surfaces of the measuring electrode (D) and 

the float body (N) with an abrasive domestic cleaning agent if 
there are deposits.

2. Clean the filter strainer (P2) under running water. Use a soft 

brush, if necessary.

3. Carefully clean all plastic parts with soap suds.
4. Use hydrochloric acid (5-10 %) for deposits.
5. Carefully rinse all parts with clean water.
6. Replace any damaged parts.

14.4 Assembling the measuring cell

• Re-assemble the measuring cell in reverse order.

– Screw in the shut-off spindles of the sample water inlet (V1) 

and outlet (V2) and the bleeding spindle (L).

– Assemble and screw in the calibration cup (I) and 

evacuation spindle (J).

– Assemble and screw in the filter cartridge (P).
– Insert the float body (N) with the sharp end facing 

downwards into the measuring tube (M).

– Tightly screw in the valve (Q2), and screw in the adjusting 

spindle (Q1).

– Fit the ascending pipe (S). Screw in the housing cylinder (T).

Fig. 25

Exploded view of screwed part / inlet chamber

Fig. 26

Exploded view of AQC-D3

Warning

Observe the safety regulations when using 
hydrochloric acid! When cleaning with 
hydrochloric acid it is absolutely necessary to 
wear safety goggles.

Caution

Before using other detergents, check the 
chemical resistance of the materials, as they may 
destroy some components.

Caution

Do not clean electrodes, filters, float bodies or 
other metal parts using hydrochloric acid.

Caution

Only tighten screw parts by hand; do not use 
tools! Risk of leaking! Ensure that all O-rings are 
fitted correctly!

TM

03

 58

92

 1

1

0

6

TM

03

 58

82

 1

1

0

6

TM

0

3

 5

902

 2

4

11

G1

D

X

W

H

W1

W1

P1

P2

P3

N

V1

O1

O

M

L

K

U1

I

G

G1

D

X

W

U2

V2

F

Q2

Q1 A1/2

S

B1

H

2

2

1

6

4

5

2

2

7

T

Summary of Contents for AquaCel AQC-D3

Page 1: ...GRUNDFOS INSTRUCTIONS AQC D3 Measuring cell Installation and operating instructions...

Page 2: ...006 A1 2009 A2 2009 EN 61000 3 3 2008 Convient uniquement aux produits avec tension de service 50 VAC ou 75 VDC IT Dichiarazione di Conformit CE Noi Grundfos dichiariamo sotto la nostra esclusiva resp...

Page 3: ...leriyle uyumlu oldu unun yaln zca bizim sorumlulu umuz alt nda oldu unu beyan ederiz D k Voltaj Y netmeli i 2006 95 EC Kullan lan standart EN 61010 1 2001 ikinci bask EMC Diretifi 2004 108 EC Kullan l...

Page 4: ...ing amplifier 15 12 5 Electrical connections 16 12 6 Checks prior to start up 17 12 7 Switching on 17 12 8 Calibrating the parameters Cl2 ClO2 O3 17 13 Operation 19 13 1 Function 19 13 2 Operation 19...

Page 5: ...English GB 5 3 Installation sketch...

Page 6: ...ll with sensor interface Fig 3 AQC D3 potentiostatic measuring cell with DIP measuring amplifier TM03 5894 2611 TM03 5895 2611 TM03 5896 2611 1 Conex DIA 1 Conex DIA 2 Conex DIA 2Q Conex DIS D measuri...

Page 7: ...c diaphragm with buffer solution PCX pH ceramic diaphragm without buffer solution PTB pH PTFE diaphragm with buffer solution PTX pH PTFE diaphragm without buffer solution PKB pH KCl filling with buffe...

Page 8: ...pressure loading valve Electrodes for disinfection AU Gold PT Platinum X No disinfection measurement Other electrodes PCB pH ceramic diaphragm with buffer solution PCX pH ceramic diaphragm without bu...

Page 9: ...armature G1 Slide I Bottom of measuring chamber K Sample water extraction device with extraction spindle L Bleeding spindle M Measuring tube N Float body O Water sensor optional O1 Screw P Filter cart...

Page 10: ...roller installed Preassembled system with Conex DIA 1 measuring amplifier and controller Measurement free chlorine Cl2 as an option with pH compensation chlorine dioxide ClO2 ozone O3 pH temperature C...

Page 11: ...electrode 10 2 1 AquaCell AQC D3 potentiostatic measuring cell Cables cable for reference electrode pH or redox single rod electrode 3 m with free cable end cable for measuring electrode 3 m with free...

Page 12: ...drilling diagram Fig 6 Drilling diagram for AQC D3 measuring cell Fig 7 Drilling diagram for preassembled systems Example with Conex DIA wall mounted unit TM03 5898 2411 300 260 360 395 121 TM03 5899...

Page 13: ...e mounting plate from factory Individual device for AQC D3 measuring cells 1 Drill four dowel holes with a diameter of 10 mm and min 60 mm deep See section 10 4 Dimensional sketches drilling diagram 2...

Page 14: ...ol output remains active The advantage of this is that in the event of frequent peaks in the sample water flow rate the alarm is not continuously triggered and the control output remains active If the...

Page 15: ...odes or electrodes A1 A2 B1 1 Cut the electrode cables to the desired length approximately 80 mm for the connections 2 Remove 80 mm of the outer insulation 3 Disentangle the braided screen until you r...

Page 16: ...mounting in a control panel Fig 19 Conex DIA 1 DIA 2 DIA 2Q for wall mounting TM03 5870 1106 Pin Component Pin 1 brown Measuring electrode M Pin 2 white Counter electrode G C Pin 3 blue Pt100 tempera...

Page 17: ...IS D If not done yet Define the basic settings for the measuring amplifier Calibrating the pH value DIP Select the right hand display for pH with the Select button Conex DIA and DIP Use the CAL button...

Page 18: ...displays with DIP The data of the last 10 calibrations can be read 12 8 3 Calibrating the parameters chlorine chlorine dioxide or ozone with Conex DIS D 1 Press the CAL selector and select the calibr...

Page 19: ...ntinuously cleaned of deposits by a cleaning wing which has a gentle yet effective action This ensures uniform sensitivity for the measuring cell over a long period The cleaning wing is driven hydraul...

Page 20: ...d for instance a gas dosing unit or dosing pump if necessary 13 4 2 After long term stop inspection If the measuring cell has been switched off for a long period of time or has been emptied for instan...

Page 21: ...these cannot be measured if this procedure is used f Incorrect photometric reference measurement during calibration Check the reference measurement and take another measurement Repeat the calibration...

Page 22: ...lectricity supply to the measuring amplifier 2 Empty the measuring cell Open the bleeding spindle L and deaeration spindle J and drain the water Unscrew the connections to the sample water inlet U1 an...

Page 23: ...re deposits 2 Clean the filter strainer P2 under running water Use a soft brush if necessary 3 Carefully clean all plastic parts with soap suds 4 Use hydrochloric acid 5 10 for deposits 5 Carefully ri...

Page 24: ...321 141 Cable for measuring electrode 3 m 96609172 45 10212 1 Water sensor with cable 1 m 96609173 45 10212 Water sensor with cable 3 m 96609174 314 605 Reference electrode 96609175 553 1561 Measuring...

Page 25: ...25...

Page 26: ...26...

Page 27: ...2 Telefax 39 02 95309290 95838461 Japan GRUNDFOS Pumps K K Gotanda Metalion Bldg 5F 5 21 15 Higashi gotanda Shiagawa ku Tokyo 141 0022 Japan Phone 81 35 448 1391 Telefax 81 35 448 9619 Korea GRUNDFOS...

Page 28: ...es it possible Innovation is the essence The name Grundfos the Grundfos logo and the payoff Be Think Innovate are registrated trademarks owned by Grundfos Management A S or Grundfos A S Denmark All ri...

Reviews: