Grundfos AquaCel AQC-D3 Installation And Operating Instructions Manual Download Page 17

English (GB)

17

Fig. 20

Conex DIS-D

Fig. 21

DIP

12.6 Checks prior to start-up

• Check the tightness of the whole system.
• Check the electrical connections.

12.7 Switching on

1. Close the bleeding spindle (L).
2. Open the shut-off spindle for sample water outlet (V2) 

2-3 turns.

3. Gradually open the shut-off spindle for sample water inlet (V2) 

2-3 turns.

– Avoid turbulent flow!

4. Check that the cleaning wing (X) rotates. 

– Try, if necessary, to briefly increase the quantity of sample 

water with the adjusting spindle (Q1) in order to move the 
cleaning wing (X).

– If the cleaning wing (X) still does not rotate, check and 

correct its position, if necessary. See section 

14. Maintenance

.

5. Switch on the measuring amplifier.
6. Only switch on the device controlled after the first calibration, 

if necessary.

12.7.1 Setting the quantity of sample water

• Set the quantity of sample water with the adjustment spindle 

(Q1) so that a small amount of sample water runs through the 
ascending pipe (S) in the housing cylinder (T). (In order for the 
water sensor to work, the float body must be at the height of 
the water sensor.)

12.8 Calibrating the parameters Cl

2

, ClO

2

, O

3

Due to the electro-chemical behaviour of the measuring cell, no 
zero-point adjustment is necessary. Only the rate of rise 
(sensitivity) must be adjusted during calibration.

• Check the calibration after 24 hours, and repeat, if necessary!

12.8.1 Photometrical measurement

1. Open the bleeding spindle (L), and let the water run for a few 

seconds.

2. Take a water sample, and close the bleeding spindle (L).
3. Determine the concentration of the measuring parameter 

photometrically, for instance using the Grundfos DIT hand 
photometer.

12.8.2 Calibration example based on chlorine measurement 

with pH compensation (not with Conex DIS-D)

• If not done yet: Define the basic settings for the measuring 

amplifier. 

Calibrating the pH value

• DIP: Select the right-hand display for pH with the Select 

button.

• Conex DIA and DIP: Use the CAL button to select the 

calibration function CAL. The LED illuminates.

1. Fill buffer solution 1 from the supply bottle into a clean vessel.
2. Immerse the combination electrode into the buffer solution.
3. If necessary, enter the code number for Cal (or full) 

authorisation.

4. Change to the menu "calibration" with the OK button.
5. Select the measured variable "pH", and confirm with OK.

– Selection possibilities: "cal. meas. value" (= measured 

value), "cal result" and "cal cycle" (= count-down function 
which triggers the alarm "Calibrate sensor" after a 
selectable time interval of 1-100 days).

TM

03 587

7 221

1

T

M

03

 58

7

8

 22

11

Pos.

Component

1

Brown

2

White

3

Black

4

Blue

5

Screen

6

Outer conductor (screen)

7

Inner conductor

8

Reference electrode

9

Measuring electrode
Counter-electrode
Pt100 temperature sensor (not with Conex DIS-C)

10

Water sensor

Caution

Observe the installation and operating 
instructions of the measuring amplifier and 
device controlled!

Caution

At first start-up or after a long-term stop: Let the 
system run for at least two hours to avoid faulty 
measurements and calibration!

7

5

1

6

9

8

2

20 2122

19

10

4

1

14 16

7

5

1

6

10

4

9

8

2

3

4

1

53 55 57

54 56 58

63 65

64 66

Caution

At first start-up or after long-term stop: Let the 
system run for at least two hours to avoid faulty 
measurements and calibration!

Caution

During the calibration: Keep the pH value, sample 
water flow rate and water temperature constant.

Caution

Observe the installation and operating 
instructions of the photometer!

Caution

Observe the installation and operating 
instructions of the measuring amplifier!

Note

When carrying out a chlorine measurement with 
pH compensation, the pH value must be 
calibrated first, since the pH value is used when 
calibrating the chlorine value!

Summary of Contents for AquaCel AQC-D3

Page 1: ...GRUNDFOS INSTRUCTIONS AQC D3 Measuring cell Installation and operating instructions...

Page 2: ...006 A1 2009 A2 2009 EN 61000 3 3 2008 Convient uniquement aux produits avec tension de service 50 VAC ou 75 VDC IT Dichiarazione di Conformit CE Noi Grundfos dichiariamo sotto la nostra esclusiva resp...

Page 3: ...leriyle uyumlu oldu unun yaln zca bizim sorumlulu umuz alt nda oldu unu beyan ederiz D k Voltaj Y netmeli i 2006 95 EC Kullan lan standart EN 61010 1 2001 ikinci bask EMC Diretifi 2004 108 EC Kullan l...

Page 4: ...ing amplifier 15 12 5 Electrical connections 16 12 6 Checks prior to start up 17 12 7 Switching on 17 12 8 Calibrating the parameters Cl2 ClO2 O3 17 13 Operation 19 13 1 Function 19 13 2 Operation 19...

Page 5: ...English GB 5 3 Installation sketch...

Page 6: ...ll with sensor interface Fig 3 AQC D3 potentiostatic measuring cell with DIP measuring amplifier TM03 5894 2611 TM03 5895 2611 TM03 5896 2611 1 Conex DIA 1 Conex DIA 2 Conex DIA 2Q Conex DIS D measuri...

Page 7: ...c diaphragm with buffer solution PCX pH ceramic diaphragm without buffer solution PTB pH PTFE diaphragm with buffer solution PTX pH PTFE diaphragm without buffer solution PKB pH KCl filling with buffe...

Page 8: ...pressure loading valve Electrodes for disinfection AU Gold PT Platinum X No disinfection measurement Other electrodes PCB pH ceramic diaphragm with buffer solution PCX pH ceramic diaphragm without bu...

Page 9: ...armature G1 Slide I Bottom of measuring chamber K Sample water extraction device with extraction spindle L Bleeding spindle M Measuring tube N Float body O Water sensor optional O1 Screw P Filter cart...

Page 10: ...roller installed Preassembled system with Conex DIA 1 measuring amplifier and controller Measurement free chlorine Cl2 as an option with pH compensation chlorine dioxide ClO2 ozone O3 pH temperature C...

Page 11: ...electrode 10 2 1 AquaCell AQC D3 potentiostatic measuring cell Cables cable for reference electrode pH or redox single rod electrode 3 m with free cable end cable for measuring electrode 3 m with free...

Page 12: ...drilling diagram Fig 6 Drilling diagram for AQC D3 measuring cell Fig 7 Drilling diagram for preassembled systems Example with Conex DIA wall mounted unit TM03 5898 2411 300 260 360 395 121 TM03 5899...

Page 13: ...e mounting plate from factory Individual device for AQC D3 measuring cells 1 Drill four dowel holes with a diameter of 10 mm and min 60 mm deep See section 10 4 Dimensional sketches drilling diagram 2...

Page 14: ...ol output remains active The advantage of this is that in the event of frequent peaks in the sample water flow rate the alarm is not continuously triggered and the control output remains active If the...

Page 15: ...odes or electrodes A1 A2 B1 1 Cut the electrode cables to the desired length approximately 80 mm for the connections 2 Remove 80 mm of the outer insulation 3 Disentangle the braided screen until you r...

Page 16: ...mounting in a control panel Fig 19 Conex DIA 1 DIA 2 DIA 2Q for wall mounting TM03 5870 1106 Pin Component Pin 1 brown Measuring electrode M Pin 2 white Counter electrode G C Pin 3 blue Pt100 tempera...

Page 17: ...IS D If not done yet Define the basic settings for the measuring amplifier Calibrating the pH value DIP Select the right hand display for pH with the Select button Conex DIA and DIP Use the CAL button...

Page 18: ...displays with DIP The data of the last 10 calibrations can be read 12 8 3 Calibrating the parameters chlorine chlorine dioxide or ozone with Conex DIS D 1 Press the CAL selector and select the calibr...

Page 19: ...ntinuously cleaned of deposits by a cleaning wing which has a gentle yet effective action This ensures uniform sensitivity for the measuring cell over a long period The cleaning wing is driven hydraul...

Page 20: ...d for instance a gas dosing unit or dosing pump if necessary 13 4 2 After long term stop inspection If the measuring cell has been switched off for a long period of time or has been emptied for instan...

Page 21: ...these cannot be measured if this procedure is used f Incorrect photometric reference measurement during calibration Check the reference measurement and take another measurement Repeat the calibration...

Page 22: ...lectricity supply to the measuring amplifier 2 Empty the measuring cell Open the bleeding spindle L and deaeration spindle J and drain the water Unscrew the connections to the sample water inlet U1 an...

Page 23: ...re deposits 2 Clean the filter strainer P2 under running water Use a soft brush if necessary 3 Carefully clean all plastic parts with soap suds 4 Use hydrochloric acid 5 10 for deposits 5 Carefully ri...

Page 24: ...321 141 Cable for measuring electrode 3 m 96609172 45 10212 1 Water sensor with cable 1 m 96609173 45 10212 Water sensor with cable 3 m 96609174 314 605 Reference electrode 96609175 553 1561 Measuring...

Page 25: ...25...

Page 26: ...26...

Page 27: ...2 Telefax 39 02 95309290 95838461 Japan GRUNDFOS Pumps K K Gotanda Metalion Bldg 5F 5 21 15 Higashi gotanda Shiagawa ku Tokyo 141 0022 Japan Phone 81 35 448 1391 Telefax 81 35 448 9619 Korea GRUNDFOS...

Page 28: ...es it possible Innovation is the essence The name Grundfos the Grundfos logo and the payoff Be Think Innovate are registrated trademarks owned by Grundfos Management A S or Grundfos A S Denmark All ri...

Reviews: