Page 5 of 22
_____________________________________________________
_____________________________________________________________________________
2.2 Operation and maintenance advice
Battery operation:
If
„
BAT
“ is shown in the lower display the battery has been used up and needs to be replaced. However, the
device will operate correctly for a certain time
If
„
BAT
“ is shown in the upper display the voltage is too low to operate the device; the battery has been
completely used up. Battery change see chapter 17 Battery change.
The battery has to be taken out, when storing device above 50°C. We recommend taking
out battery if device is not used for a longer period.
After recommissioning, the real-time clock has to be set again.
Use device and sensor carefully and according to its technical data (do not throw it, strike
it, etc.). Protect the device from dirt.
USB or mains operation:
When connecting a mains cable or USB interface cable, please take care to connect only
allowed components.
The output voltage of a connected power supply unit has to be between 4.5 and 5.5 V DC.
Do not apply overvoltage!
We recommend operation with interface cable USB 5100. Then device is supplied by the
USB interface of the connected PC or USB power supply adapter.
3
Start of operation
Connect sensor, switch instrument on with
….
After the segment test
the instrument shows
“
[ORR
“ shortly, if it was user adjusted.
The device starts measurement afterwards.
4
Operation
4.1 Display elements
1
Main display
:
Oxygen concentration in % (% Vol. O
2
) or
Oxygen partial pressure (hPa or. mmHg)
Choice via
- key
2
Secondary display
:
sensor temperature or absolute pressure, (alternating,
please refer to chapter 5 Configuration
“
LCD.2
“
3
Main display units
4
State of battery
5
Shows, if minimum/maximum/
memorized measuring value is in display
6
Arrow
“ok“: Measured value is stable
7
Arrow
„CAL“: Calibration
8
Arrow
“Logg“: Logger function is chosen
Is blinking, if cyclic logger is running
9
Arrow
“hPa“: Pressure unit of internal sensor